心室辅助装置导线生物膜形成与迁移研究
本研究通过实验分析HeartWare HVAD导线表面化学与生物膜形成,采用滴流生物膜反应器模拟出口部位、隧道式间质生物膜试验模拟皮下组织,并检查临床取出的导线。结果显示常见病原体可黏附并形成生物膜,葡萄球菌在隧道模型中生物膜形成量比出口模型高500至10000倍,导线绒面三维结构和硅胶粘合剂促进生物膜生长,临床导线绒面存在微间隙。
要点
- 常见导线感染病原体可黏附HVAD导线光滑和绒面部分并形成生物膜。
- 葡萄球菌在隧道模型中的生物膜形成量比出口模型高500至10000倍。
- 导线绒面三维结构和硅胶粘合剂促进生物膜生长,临床导线存在微间隙。
核心结论
- 病原体对导线不同部位有偏好,皮下隧道对生物膜形成和迁移重要,临床导线微间隙可能促进侵袭性感染。
研究设计与样本
- 实验分析HeartWare HVAD导线表面化学和生物膜形成,用滴流生物膜反应器模拟出口部位,用隧道式间质生物膜试验模拟皮下组织,并用扫描电镜检查心脏移植时取出的临床HVAD导线。
主要发现
- 常见导线感染病原体可黏附HVAD导线光滑和绒面部分并在滴流反应器中形成稳固生物膜,其中铜绿假单胞菌和白色念珠菌生物量更大。
- 表皮葡萄球菌、金黄色葡萄球菌和白色念珠菌在间质隧道中显示生物膜迁移,而铜绿假单胞菌无迁移。
- 葡萄球菌在隧道模型中的生物膜形成量比出口模型高500至10000倍。
- 导线绒面三维结构和硅胶粘合剂促进生物膜生长,临床取出导线绒面全程组织长入不足且纤维间存在微间隙。
临床怎么用
- 应关注导线出口部位和皮下隧道不同部位的感染风险,针对葡萄球菌和真菌等病原体加强监测与预防。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Driveline infections remain an important complication of ventricular assist device therapy, with biofilm formation being a major contributor. This study aimed to elucidate factors that govern biofilm formation and migration on clinically relevant ventricular assist device drivelines. METHODS: Experimental analyses were performed on HeartWare HVAD (HeartWare International Inc, Framingham, Mass) drivelines to assess surface chemistry and biofilm formation. To mimic the driveline exit site, a drip-flow biofilm reactor assay was used. To mimic a subcutaneous tissue environment, a tunnel-based interstitial biofilm assay was developed. Clinical HVAD drivelines explanted at the time of cardiac transplantation were also examined by scanning electron microscopy. RESULTS: Common causative pathogens of driveline infections were able to adhere to the smooth and velour sections of the HVAD driveline and formed robust biofilms in the drip-flow biofilm reactor; however, Pseudomonas aeruginosa and Candida albicans had greater biomass. Biofilm migration within the interstitial driveline tunnel was evident for Staphylococcus epidermidis, Staphylococcus aureus, and C albicans, but not P aeruginosa. Biofilm formation by staphylococci was 500 to 10,000 times higher in the tunnel-based model compared with our exit site model. The 3-dimensional structure of the driveline velour and the use of silicone adhesive in driveline manufacturing were found to promote biofilm growth, and explanted patient drivelines demonstrated inadequate tissue in-growth along the entire velour with micro-gaps between velour fibers. CONCLUSIONS: This work highlights the predilection of pathogens to different parts of the driveline, the importance of the subcutaneous tunnel to biofilm formation and migration, and the presence of micro-gaps in clinical drivelines that could facilitate invasive driveline infections.
引用
- Qu Y, McGiffin D, Kure C, et al. Biofilm formation and migration on ventricular assist device drivelines. J Thorac Cardiovasc Surg. 2019. PMID: 30955967. doi:10.1016/j.jtcvs.2019.02.088
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。