腔肺辅助装置计算流体动力学模拟用于衰竭Fontan循环

Computational fluid dynamic simulations of a cavopulmonary assist device for failing Fontan circulation.
PubMed 文献 · J Thorac Cardiovasc Surg · 2019-03-21

研究用计算流体动力学模拟1个理想化与2个患者特异性心外Fontan几何模型,评估多腔导管联合体外血泵的腔肺辅助性能。结果显示理想化与患者特异性模型分别获得最高4与9 mm Hg的压力增益,溶血指标低于2%阈值,壁面剪切应力在肺血管与导管局部升高。

要点

  • 多腔导管联合体外血泵在模拟中可提供最高4与9 mm Hg压力增益
  • 高泵 intake 流量下溶血指标低于2%阈值数倍
  • 壁面剪切应力在肺血管与导管装置区域升高

核心结论

  • 多腔导管在模拟中显示作为经皮桥接支持部分衰竭心外Fontan患者的潜力,局限将在后续设计与实验中处理。

研究设计与样本

  • 对1个理想化与2个患者特异性心外Fontan几何模型进行计算流体动力学模拟,模拟范围涵盖生理流量与泵辅助水平。

主要发现

  • 理想化与患者特异性模型分别获得最高4与9 mm Hg压力增益,压力增益随心输出量增高与体外泵 intake 增大而增加。
  • 流量加权溶血显示最高泵 intake 流量病例的血红蛋白损伤水平低于2%阈值数倍,壁面剪切应力在肺血管与导管装置区域升高。

关键数值与推荐

  • 压力增益最高4 mm Hg(理想化模型)与9 mm Hg(患者特异性模型);溶血指标低于2%阈值数倍(最高泵 intake 流量病例)。

临床怎么用

  • 该装置有望作为经皮桥接支持用于部分衰竭心外Fontan患者,但需进一步设计与实验验证。

局限性

  • 研究为计算流体动力学模拟,仅含1个理想化与2个患者特异性模型,尚未经实验或临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Adult patients who have undergone the Fontan procedure are highly vulnerable to gradual, progressive circulatory failure, and options to reverse this situation are few. A cavopulmonary assist device could decongest the venous and lymphatic systems, overcome elevated pulmonary vascular resistance, increase cardiac output, and support some of these patients to heart transplant. This study characterizes the performance and challenges of a novel multilumen cannula coupled to an external blood pump proposed as a potential Fontan cavopulmonary assist strategy. METHODS: Computational fluid dynamic simulations were conducted for 3 extracardiac Fontan geometries consisting of 1 idealized model and 2 patient-specific models. A range of physiologic flow rates and pump assist levels were simulated to calculate the pressure gain provided by the multilumen cannula. Hemolysis index was estimated for the idealized model with Lagrangian particle tracking and 2 variations of the power-law. Wall shear stresses were also examined. RESULTS: Pressure gains up to 4 and 9 mm Hg were achieved for the idealized and patient-specific models, respectively. Pressure gains increased with both higher cardiac output and larger pump intake through the external pump. Flow-weighted hemolysis show hemoglobin damage levels to be several times lower than the 2% threshold at the highest pump intake flow cases. Wall shear stress predictions depict elevated areas in the pulmonary vessels and regions of the cannula device. CONCLUSIONS: The cannula tested in this study shows promise as a percutaneous option to bridge support in some patients with a failing extracardiac Fontan. Limitations identified will be addressed in future design iterations and in ongoing experimental tests.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Lin WCP, Doyle MG, Roche SL, et al. Computational fluid dynamic simulations of a cavopulmonary assist device for failing Fontan circulation. J Thorac Cardiovasc Surg. 2019. PMID: 31005303. doi:10.1016/j.jtcvs.2019.03.008