主动脉瓣修复术中测试装置

An intraoperative test device for aortic valve repair.
PubMed 文献 · J Thorac Cardiovasc Surg · 2018-10-24

该研究开发了一种可在术中对主动脉瓣进行测试和可视化的装置,用于提高主动脉瓣修复的可重复性。装置可固定在横断的主动脉内,使主动脉根部加压并在内镜下观察闭合的瓣膜。在离体猪心中演示了该装置用于观察主动脉瓣及测量瓣叶对合高度,操作时间少于1分钟。

要点

  • 装置可在术中固定于横断主动脉,加压后内镜下观察闭合瓣膜。
  • 操作步骤(置入固定、加压、内镜观察)少于1分钟。
  • 可用于修复前评估反流机制及修复后测试瓣膜闭合功能。

核心结论

  • 该装置能在生理压力下更相关地术前评估瓣膜并测试修复后瓣膜闭合功能。

研究设计与样本

  • 研究开发了一种术中测试装置,并在离体猪心中演示其用于观察主动脉瓣和测量瓣叶对合高度。

主要发现

  • 装置可容易且可逆地附着于主动脉根部,在模拟舒张期条件下直接检查主动脉瓣。
  • 操作步骤包括置入固定装置、加压瓣膜和内镜观察闭合瓣膜,完成时间少于1分钟。
  • 装置可用于测量舒张状态下瓣叶对合高度。

临床怎么用

  • 外科医生可在修复前立即术中研究瓣膜以确定功能不全机制,修复后立即评估闭合功能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Aortic valve repair is currently in transition from surgical improvisation to a reproducible operation and an option for many patients with aortic regurgitation. Our research efforts at improving reproducibility include development of methods for intraoperatively testing and visualizing the valve in its diastolic state. METHODS: We developed a device that can be intraoperatively secured in the transected aorta allowing the aortic root to be pressurized and the closed valve to be inspected endoscopically. Our device includes a chamber that can be pressurized with crystalloid solution and ports for introduction of an endoscope and measuring gauges. We show use of the device in explanted porcine hearts to visualize the aortic valve and to measure leaflet coaptation height in normal valves and in valves that have undergone valve repair procedures. RESULTS: The procedure of introducing and securing the device in the aorta, pressurizing the valve, and endoscopically visualizing the closed valve is done in less than 1 minute. The device easily and reversibly attaches to the aortic root and allows direct inspection of the aortic valve under conditions that mimic diastole. It enables the surgeon to intraoperatively study the valve immediately before repair to determine mechanisms of incompetence and immediately after the repair to assess competence. We also show its use in measuring valve leaflet coaptation height in the diastolic state. CONCLUSIONS: This device enables more relevant prerepair valve assessment and also enables a test of postrepair valve competence under physiological pressures.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。