文献速递瓣膜外科

高保真微创二尖瓣手术模拟器的研制

Development of a high-fidelity minimally invasive mitral valve surgery simulator.
PubMed 文献 · J Thorac Cardiovasc Surg · 2018-09-29

本研究基于经验丰富二尖瓣外科医生的访谈需求,采用工业系列设计流程研制了支持内镜与机器人入路的高保真微创二尖瓣手术模拟器,含一次性硅胶二尖瓣装置与基于4摄像头及边缘检测算法的反馈系统。反馈系统可测量缝线宽度与深度,深度与显微CT测量无显著差异(P = .139),宽度存在显著差异(P = .001)并经回归方程校准后无显著差异(P = .865)。99名外科医生试用后多数认同该模拟器是训练微创二尖瓣手术的良好方法。

要点

  • 模拟器支持内镜和机器人入路的微创二尖瓣手术训练。
  • 反馈系统基于4摄像头与边缘检测算法测量缝线深度和宽度。
  • 深度测量与显微CT一致,宽度经校准后一致。

核心结论

  • 成功研制了面向内镜与机器人入路的高保真微创二尖瓣手术模拟器,可客观、可重复地训练相关技能。

研究设计与样本

  • 基于访谈经验丰富二尖瓣外科医生获得的预设需求,应用工业系列设计流程开发模拟器。
  • 模拟器包含带内镜与机器人入路的胸廓躯干模型,以及带反馈系统的一次性硅胶二尖瓣装置。
  • 反馈系统基于硅胶瓣膜周围4个摄像头与边缘检测算法计算缝线深度和宽度。
  • 通过将模拟器测量值与同一缝线的三维重建显微CT扫描手动测量值比较来评估测量效度。
  • 2014至2018年间由独立外科医生试用模拟器,并通过问卷进行评估。

主要发现

  • 深度在模拟器与显微CT测量之间无显著差异(P = .139)。
  • 宽度在模拟器与显微CT测量之间存在显著差异(P = .001),随后找到显著回归方程(P < .0001)用于校准模拟器。
  • 校准后宽度测量无显著差异(P = .865)。
  • 共99名外科医生试用模拟器,多数认同其为训练微创二尖瓣手术的良好方法,且二尖瓣与缝线放置的外观和手感逼真。

关键数值与推荐

  • 深度比较P = .139;宽度比较P = .001;校准回归方程P < .0001;校准后宽度比较P = .865。
  • 99名外科医生参与试用与问卷评估。

临床怎么用

  • 该模拟器可用于内镜与机器人微创二尖瓣手术的技能训练,并以客观、可重复的方式提供缝线深度与宽度反馈。
  • 作者指出仍需未来研究为其在外科医生培训中的应用提供证据。

局限性

  • 原文未报告

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The aim of this study was to develop a high-fidelity minimally invasive mitral valve surgery (MIMVS) simulator. METHODS: The process of industrial serial design was applied based on pre-set requirements, acquired by interviewing experienced mitral surgeons. A thoracic torso with endoscopic and robotic access and disposable silicone mitral valve apparatus with a feedback system was developed. The feedback system was based on 4 cameras around the silicone valve and an edge detection algorithm to calculate suture depth and width. Validity of simulator measurements was assessed by comparing simulator-generated values with measurements done manually on 3-dimensional reconstructed micro-computed tomography scan of the same sutures. Independent surgeons tested the simulator between 2014 and 2018, whereupon an evaluation was done through a questionnaire. RESULTS: The feedback system was able to provide width and depth measurements, which were subsequently scored by comparison to pre-set target values. Depth did not significantly differ between simulator and micro-computed tomography scan measurements (P = .139). Width differed significantly (P = .001), whereupon a significant regression equation was found (P < .0001) to calibrate the simulator. After calibration, no significant difference was found (P = .865). In total, 99 surgeons tested the simulator and more than agreed with the statements that the simulator is a good method for training MIMVS, and that the mitral valve and suture placement looked and felt realistic. CONCLUSIONS: We successfully developed a high-fidelity MIMVS simulator for endoscopic and robotic approaches. The simulator provides a platform to train skills in an objective and reproducible manner. Future studies are needed to provide evidence for its application in training surgeons.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Sardari Nia P, Daemen JHT, Maessen JG, et al. Development of a high-fidelity minimally invasive mitral valve surgery simulator. J Thorac Cardiovasc Surg. 2018. PMID: 30385017. doi:10.1016/j.jtcvs.2018.09.014

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。