猪二尖瓣连枷修复后6个月:ePTFE多弦装置性能与愈合

Performance and healing of an expanded polytetrafluoroethylene multichordal device at 6 months after repair of mitral leaflet flail in swine.
PubMed 文献 · J Thorac Cardiovasc Surg · 2018-09-25

研究在7只猪中经二尖瓣腱索离断建立二尖瓣脱垂模型,用含4根新腱索的ePTFE多弦装置(MitraPatch)修复,6只存活至150或180天并每月超声随访。术后反流显著减轻,装置完整且宿主反应可控,但1只出现装置撕脱。

要点

  • 装置在多弦二尖瓣脱垂猪模型中6个月血流动力学良好
  • 装置完整无撕裂、拉伸或纤维化,宿主反应可控
  • 1只猪因误断交界与支柱腱索出现装置撕脱和1-2+反流

核心结论

  • 多弦装置设计的外科解剖原则在猪6个月模型中显示有效,血流动力学表现良好且宿主反应可控。

研究设计与样本

  • 7只猪经二尖瓣腱索离断建立脱垂模型,1只用于手术开发,6只植入装置并存活至150天(n=4)或180天(n=2),每月超声随访。

主要发现

  • 腱索离断后所有猪均出现重度二尖瓣反流。
  • 术后即刻反流降至无或微量3只、1级3只、1至2+级1只。
  • 无或微量及1级反流的猪中装置完整、对合完全恢复,终止时装置无撕脱、拉伸或纤维化。
  • 装置的新瓣叶与新乳头肌部分引发宿主反应,有望形成活体瓣膜替代组织,但远期可控性未知。
  • 1至2+级反流的猪死亡时见该区域交界与支柱腱索意外离断,装置在瓣叶插入处撕脱。

临床怎么用

  • 该装置在猪模型中6个月表现支持进一步开展临床前研究,但需关注瓣叶插入处撕脱风险。

局限性

  • 样本量小(n=7),仅猪模型,随访6个月,无法判断宿主反应多年后的可控性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Repairing a prolapsed mitral valve that involves multiple cusps is procedurally complicated and carries a higher risk of failure when performed with individual neochordae. Inserting multiple neochordae into the papillary muscle, sizing, and aligning them in a manner that can restore coaptation is challenging. A multichordal mitral valve apparatus (MitraPatch) with a single neopapillary muscle section, 4 neochordae with each chord ending in a neoleaflet section, was developed. In this study, the 6-month outcomes of using this apparatus to repair mitral prolapse in swine is reported. METHODS: Seven pigs (n = 7) with prolapse from mitral chordal transection were repaired with the device. One pig was used to develop the procedure, and 6 swine received the implant and were survived to 150 days (n = 4) and 180 days (n = 2), with monthly echocardiographic examinations. All hearts were explanted for gross necropsy and detailed histopathology. RESULTS: Severe mitral regurgitation (MR) was observed after chordal transection in all pigs. Repairing the valve with the device reduced MR to none-or-trace levels in 3 swine, grade 1 in 3 swine, and 1 to 2+ in 1 swine immediately after surgery. In the pigs with none/trace/grade 1 MR, the device was intact and coaptation was fully restored as observed on serial echocardiograms. The device was intact, without dehiscence, stretching, or fibrosis at termination. The neoleaflet and neopapillary sections of the device elicited a host response, which is on track to produce living valve replacement tissue, but we cannot tell how controlled this might be several years later. In the pig with grade 1 to 2+ MR, accidental transection of both marginal and strut chordae in that region was observed at death, with dehiscence of the device at the site of its insertion into the leaflet. CONCLUSIONS: The anatomic principles that guide this multichordal design appear to be valid, with good hemodynamic performance and a controlled host response at 6 months.