患者特异性形态对经导管主动脉瓣置换术后窦部血流淤滞的影响:一项体外研究
该体外研究选取两例代表性患者主动脉根进行三维打印并植入CoreValve或SAPIEN瓣膜,用高时空分辨率粒子图像测速评估窦部血流。SAPIEN瓣膜窦部峰值速度约为CoreValve的3倍(P<.01),涡量大小在患者模型间差异显著(P<.01),窦部冲刷特征与患者和瓣膜类型均强相关。
要点
- SAPIEN瓣膜窦部峰值速度约为CoreValve的3倍(P<.01)。
- 不同患者模型间涡量大小差异显著(P<.01),与瓣膜类型无关。
- 窦部血流动力学对主动脉根特征及瓣膜-根交互高度敏感。
核心结论
- 窦部血流动力学对主动脉根特征及经导管主动脉瓣与主动脉根的交互高度敏感。
- 代表性患者模型间窦部血流冲刷和淤滞区域的差异可能反映不同的瓣叶血栓形成或瓣膜退变风险。
研究设计与样本
- 体外研究:选取两例代表性患者主动脉根,分割并三维打印后植入Medtronic CoreValve和Edwards SAPIEN经导管主动脉瓣。
- 采用高时空分辨率粒子图像测速评估窦部血流动力学,进行窦部涡旋追踪、剪切应力概率密度函数和窦部冲刷分析,作为瓣膜类型和患者形态的函数进行独立病例评估。
主要发现
- SAPIEN瓣膜窦部峰值速度约为CoreValve的3倍(两模型分别为0.30 ± 0.02 m/s和0.34 ± 0.041 m/s vs 0.13 ± 0.01 m/s和0.10 ± 0.02 m/s)(P < .01)。
- 不同代表性患者模型间涡量大小差异显著(75 ± 1.1 s-1、77 ± 3.2 s-1、109 ± 2.3 s-1和250 ± 4.1 s-1)(P < .01),与瓣膜类型无关。
- 窦部血液冲刷特征随心动周期的变化与患者和瓣膜类型均强相关。
- 较小的窦部和窦管交界处使流体动力学倾向于有利的剪切应力和冲刷特征,SAPIEN瓣膜进一步放大了这一效应。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The goal of this study is to evaluate how sinus flow patterns after transcatheter aortic valve replacement in realistic representative patient roots vary. Sinus flow can affect transcatheter aortic valve operation and likely leaflet thrombosis occurrence due to stasis and poor washout. How the interaction between transcatheter aortic valve and representative patient aortic roots affects sinus hemodynamics is important to establish for future individualization of transcatheter aortic valve replacement therapy. METHODS: Two representative patient aortic roots were selected, segmented and 3-dimensional printed followed by deployment of Medtronic CoreValve (Medtronic Inc, Irvine, Calif) and Edwards SAPIEN (Edwards Lifesciences, Irvine Calif) transcatheter aortic valves. Sinus hemodynamics were assessed in vitro using high spatio-temporal resolution particle-image-velocimetry. Detailed sinus vortex tracking, shear stress probability density functions, and sinus washout were evaluated and assessed as a function of valve type and representative patient morphology as independent case studies. RESULTS: Peak velocity in the sinus with SAPIEN valve was approximately 3 times higher than with CoreValve for both models (0.30 ± 0.02 m/s and 0.34 ± 0.041 m/s vs 0.13 ± 0.01 m/s and 0.10 ± 0.02 m/s) (P < .01). Between representative patient models, vorticity magnitudes were significantly different (75 ± 1.1 s-1, 77 ± 3.2 s-1, 109 ± 2.3 s-1, and 250 ± 4.1 s-1) (P < .01) regardless of valve type. Sinus blood washout characteristic as a function of cardiac cycles was strongly both patient related and valve specific. Fluid dynamics favored shear stresses and washout characteristics due to a smaller sinus and sinotubular junction, further amplified by the SAPIEN valve. CONCLUSIONS: Sinus flow dynamics are highly sensitive to aortic root characteristics and transcatheter aortic valve aortic root interaction. Differences in sinus-flow washout and stasis regions between representative patient models may be reflected in different risks of leaflet thrombosis or valve degeneration.