PubMed 文献 · J Thorac Cardiovasc Surg · 2018-05-28

本研究对26毫米Edwards SAPIEN 3经导管主动脉瓣进行高分辨率微型CT扫描并建立三维几何网格,采用有限元分析模拟体循环压力加载,计算支架与瓣叶应力。结果显示舒张期瓣叶最大主应力为2.7 MPa、最小主应力为-0.47 MPa,支架在80 mm Hg下最大主应力为38.2 MPa、最小主应力为-44.4 MPa,峰值应力均位于瓣叶与支架连接的连合处;Dacron裙边对瓣叶和支架应力影响极小。

要点

  • 26毫米SAPIEN 3瓣膜舒张期瓣叶最大主应力为2.7 MPa,最小主应力为-0.47 MPa。
  • 80 mm Hg下支架最大主应力为38.2 MPa,最小主应力为-44.4 MPa。
  • 瓣叶与支架峰值应力均位于连合处,该区域最可能起始退变。

核心结论

  • 26毫米SAPIEN 3瓣膜的支架和瓣叶峰值应力均位于瓣叶附着的连合尖端,这些区域最可能起始退变。
  • Dacron裙边对瓣叶和支架应力的影响极小。

研究设计与样本

  • 对商用26毫米Edwards SAPIEN 3瓣膜进行高分辨率微型CT扫描,建立支架和瓣膜的精确三维几何网格。
  • 瓣叶材料属性取自外科生物瓣,支架材料属性基于钴铬合金。
  • 模拟体循环压力加载,采用有限元分析计算应力。

主要发现

  • 舒张期经导管主动脉瓣瓣叶最大主应力为2.7 MPa,最小主应力为-0.47 MPa。
  • 瓣叶峰值应力见于瓣叶连合处上部与支架连接处。
  • 80 mm Hg下支架最大主应力为38.2 MPa,最小主应力为-44.4 MPa,位于连合支架下方。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Transcatheter aortic valve replacement has proven successful in treating intermediate-risk, high-risk, and inoperable patients with severe aortic stenosis. Third-generation, balloon-expandable transcatheter aortic valves were developed with an outer sealing skirt to reduce paravalvular leakage. As transcatheter aortic valve replacement use expands, long-term durability questions remain. Valve design influences durability, where regions of increased leaflet stress are vulnerable to early degeneration. However, third-generation transcatheter aortic valve stresses are unknown. Our goals were to determine the stent and leaflet stresses of third-generation, balloon-expandable transcatheter aortic valves. METHODS: The commercial 26-mm Edwards SAPIEN 3 valve (Edwards Lifesciences, Inc, Irvine, Calif) underwent high-resolution micro-computed tomography scanning to develop a precise 3-dimensional geometric mesh of the stent and valve. Leaflet material properties were obtained from surgical bioprostheses, and stent material properties were based on cobalt-chromium. Simulations of systemic pressure loading were performed, and stress was calculated using finite element analyses. RESULTS: At diastole, maximum and minimum principal stresses on transcatheter aortic valve leaflets were 2.7 MPa and -0.47 MPa, respectively. Peak leaflet stresses were observed at upper leaflet commissures, at their connection to the stent. Maximum and minimum principal stresses for the stent were 38.2 MPa and -44.4 MPa, respectively, at 80 mm Hg and were located just below the commissural stent. CONCLUSIONS: Stress analysis of the 26-mm SAPIEN 3 valve using exact geometry from high-resolution scans demonstrated that peak stresses for both transcatheter aortic valve stent and leaflets were present at commissural tips where leaflets were attached. These regions would be most likely to initiate degeneration. The Dacron skirt had minimal effect on stresses on leaflets and stent.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Xuan Y, Dvir D, Wang Z, et al. Stent and leaflet stresses in 26-mm, third-generation, balloon-expandable transcatheter aortic valve. J Thorac Cardiovasc Surg. 2018. PMID: 30041923. doi:10.1016/j.jtcvs.2018.04.115

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。