儿童连续流全人工心脏的早期在体经验
本研究在4只羔羊(平均体重28.7±2.3 kg)体内植入儿童连续流全人工心脏(P-CFTAH)原型,经正中开胸在体外循环下手术,评估其血流动力学性能。结果显示解剖匹配良好、植入容易,平均主动脉阻断时间98±18分钟,基线血流动力学稳定,左右心房压差在体循环与肺血管阻力比值0.7至12范围内维持于±5 mm Hg以内,并成功实现泵速调制产生动脉搏动。结论认为该初期原型满足自调节、性能和脉冲调制的要求。
要点
- P-CFTAH原型在4只羔羊体内植入,解剖匹配良好且植入容易。
- 平均主动脉阻断时间98±18分钟,泵速3.4±0.2 krpm,泵流量2.1±0.9 L/min。
- 左右心房压差在体循环与肺血管阻力比值0.7至12范围内维持于±5 mm Hg以内。
核心结论
- 该初期P-CFTAH原型满足自调节、性能和脉冲调制的既定要求。
研究设计与样本
- 急性在体研究,将P-CFTAH植入4只羔羊(平均体重28.7±2.3 kg),经正中开胸在体外循环下手术,记录肺循环和体循环泵性能参数。
主要发现
- 实验显示解剖匹配良好且植入容易,平均主动脉阻断时间为98±18分钟。
- 全部4只动物基线血流动力学稳定:泵速3.4±0.2 krpm,泵流量2.1±0.9 L/min,功率3.0±0.8 W,动脉压68±10 mm Hg,左右心房压均为6±1 mm Hg。
- 在体循环与肺血管阻力比值0.7至12的广泛范围内,无论是否进行泵速调制,左右心房压差均维持于预期限值±5 mm Hg以内。
- 成功进行泵速调制以产生动脉搏动。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Heart transplantation in infants and children is an accepted therapy for end-stage heart failure, but donor organ availability is low and always uncertain. Mechanical circulatory support is another standard option, but there is a lack of intracorporeal devices due to size and functional range. The purpose of this study was to evaluate the in vivo performance of our initial prototype of a pediatric continuous-flow total artificial heart (P-CFTAH), comprising a dual pump with one motor and one rotating assembly, supported by a hydrodynamic bearing. METHODS: In acute studies, the P-CFTAH was implanted in 4 lambs (average weight: 28.7 ± 2.3 kg) via a median sternotomy under cardiopulmonary bypass. Pulmonary and systemic pump performance parameters were recorded. RESULTS: The experiments showed good anatomical fit and easy implantation, with an average aortic cross-clamp time of 98 ± 18 minutes. Baseline hemodynamics were stable in all 4 animals (pump speed: 3.4 ± 0.2 krpm; pump flow: 2.1 ± 0.9 liters/min; power: 3.0 ± 0.8 W; arterial pressure: 68 ± 10 mm Hg; left and right atrial pressures: 6 ± 1 mm Hg, for both). Any differences between left and right atrial pressures were maintained within the intended limit of ±5 mm Hg over a wide range of ratios of systemic-to-pulmonary vascular resistance (0.7 to 12), with and without pump-speed modulation. Pump-speed modulation was successfully performed to create arterial pulsation. CONCLUSION: This initial P-CFTAH prototype met the proposed requirements for self-regulation, performance, and pulse modulation.
引用
- Karimov JH, Horvath DJ, Byram N, et al. Early in vivo experience with the pediatric continuous-flow total artificial heart. J Heart Lung Transplant. 2018. PMID: 29703578. doi:10.1016/j.healun.2018.03.019
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。