用于Fontan患者的自驱动主动脉涡轮静脉辅助装置的体外验证
本研究开发了一种无需外部驱动电源的全植入式整合主动脉涡轮静脉辅助(iATVA)系统,并在脉动体外循环中验证单心室生理后,测量其在小儿和成人生理条件下的血流动力学性能。原型装置以约800转/分运行,最多从主动脉抽取10%体循环血液驱动血液涡轮,被动驱动混流静脉泵,在供应全部腔静脉血流的同时产生约5 mm Hg的静脉压恢复。
要点
- iATVA系统以约800转/分运行,最多从主动脉抽取10%体循环血液。
- 该系统在供应全部腔静脉血流时产生约5 mm Hg静脉压恢复。
- 全植入式iATVA系统在体外实验中可行,无需外部电源。
核心结论
- 实验表明全植入式iATVA系统可行,可消除Fontan机械静脉辅助对外部电源的需求,并对抗术后静脉重塑和Fontan衰竭。
研究设计与样本
- 体外研究:开发共旋转涡轮和泵叶轮叶片设计并制造3个原型,在脉动体外回路中验证单心室生理后测量iATVA系统的血流动力学性能。
主要发现
- 原型iATVA装置以约800转/分运行,最多从主动脉抽取10%体循环血液,利用其流体动力驱动血液涡轮,进而被动驱动混流静脉泵。
- 通过将单心室出口的部分可用能量转移至静脉侧,iATVA系统在供应全部腔静脉血流时产生约5 mm Hg的静脉压恢复。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Palliative repair of single ventricle defects involve a series of open-heart surgeries where a single-ventricle (Fontan) circulation is established. As the patient ages, this paradoxical circulation gradually fails, because of its high venous pressure levels. Reversal of the Fontan paradox requires an extra subpulmonic energy that can be provided through mechanical assist devices. The objective of this study was to evaluate the hemodynamic performance of a totally implantable integrated aortic-turbine venous-assist (iATVA) system, which does not need an external drive power and maintains low venous pressure chronically, for the Fontan circulation. METHODS: Blade designs of the co-rotating turbine and pump impellers were developed and 3 prototypes were manufactured. After verifying the single-ventricle physiology at a pulsatile in vitro circuit, the hemodynamic performance of the iATVA system was measured for pediatric and adult physiology, varying the aortic steal percentage and circuit configurations. The iATVA system was also tested at clinical off-design scenarios. RESULTS: The prototype iATVA devices operate at approximately 800 revolutions per minute and extract up to 10% systemic blood from the aorta to use this hydrodynamic energy to drive a blood turbine, which in turn drives a mixed-flow venous pump passively. By transferring part of the available energy from the single-ventricle outlet to the venous side, the iATVA system is able to generate up to approximately 5 mm Hg venous recovery while supplying the entire caval flow. CONCLUSIONS: Our experiments show that a totally implantable iATVA system is feasible, which will eliminate the need for external power for Fontan mechanical venous assist and combat gradual postoperative venous remodeling and Fontan failure.
引用
- Pekkan K, Aka IB, Tutsak E, et al. In vitro validation of a self-driving aortic-turbine venous-assist device for Fontan patients. J Thorac Cardiovasc Surg. 2018. PMID: 29666009. doi:10.1016/j.jtcvs.2018.02.088