术中旁路移植血管血流测量:传输时间血流仪(TTFM)的临床评估
本研究在60例使用左乳内动脉(LIMA)作为桥血管的冠状动脉旁路移植术(CABG)患者中,对比TTFM与游离血流测量,发现TTFM总体高估7.1%±16.3%,4mm探头及高流量是独立决定因素,最佳高估阈值为68 mL/min。
要点
- TTFM在64%的测量中高于游离血流,总体高估7.1%±16.3%。
- 4mm探头测量高估13.3%±15.4%,探头过大(OR 9.56)和高流量(OR 1.02)为独立决定因素。
- TTFM高估的最佳阈值为68 mL/min,总体精度临床可接受。
核心结论
- TTFM总体高估7.1%,是CABG术中LIMA血流的准确指标,精度临床可接受。
- 流量大于68 mL/min或使用过大探头时可能出现高估。
研究设计与样本
- 观察性研究,纳入60例使用LIMA作为桥血管的CABG患者,基于120对术中测量(血管扩张前后)。
主要发现
- TTFM在64%的测量中高于游离血流,总体高估7.1%±16.3%(扩张前)。
- 4mm探头测量高估13.3%±15.4%(p<0.01)。
- 多因素回归显示探头过大(OR 9.56,95%CI 2.03-45.10,p=0.004)和高流量(OR 1.02,95%CI 1.01-1.04,p<0.001)是TTFM高估的独立决定因素。
- Bland-Altman分析显示血管扩张后无系统性高估。
- ROC分析确定高估的最佳阈值为68 mL/min。
临床怎么用
- 指南倡导CABG术中采用TTFM验证桥血管通畅性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Guidelines advocate transit time flowmetry (TTFM) for intraoperative graft patency verification during coronary artery bypass graft surgery (CABG), but studies on accuracy and precision of the TTFM technique are few. In an observational study of CABG patients, we analyzed covariation of left internal mammary artery (LIMA) blood flow with TTFM and free flow measurements. METHODS: Covariation of TTFM and free blood flows was evaluated in 60 patients undergoing CABG using the LIMA as one of the conduits. With LIMA flow measurements routinely performed before and after vasodilation, results are based on 120 paired intraoperative measurements. RESULTS: As demonstrated by a combined approach of regression and Bland-Altman analysis for the two flow situations, TTFM was higher than free flow in 64% of measurements, with an overestimation by TTFM of 7.1% ± 16.3% in the overall cohort (prevasodilation), statistically carried by measurements with 4-mm probes (overestimation by 13.3% ± 15.4%, both p < 0.01). In a multiregression analysis, oversizing of the TTFM probe (odds ratio 9.56, 95% confidence interval: 2.03 to 45.10, p = 0.004) and high flows (odds ratio 1.02, 95% confidence interval: 1.01 to 1.04, p < 0.001) were independent determinants of flow overestimation by TTFM, although in the Bland-Altman analysis no systematic overestimation was seen in the postvasodilation situation. In a receiver-operating characteristics analysis, optimal cutoff value as determined from Youden's index for assuming flow overestimation was 68 mL/min. CONCLUSIONS: Overall, with slight overall overestimation of 7.1%, TTFM is an accurate indicator of LIMA blood flow during CABG, with a clinically acceptable precision. Overestimation may be expected with flows greater than 68 mL/min, but most importantly, in situations with oversized TTFM probes.
引用
- Amin S, Werner RS, Madsen PL, et al. Intraoperative Bypass Graft Flow Measurement With Transit Time Flowmetry: A Clinical Assessment. Ann Thorac Surg. 2018. PMID: 29605596. doi:10.1016/j.athoracsur.2018.02.067
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。