文献速递瓣膜外科

一种旨在简化人工腱索长度测定的新型二尖瓣成形环的初步应用经验

Initial Experience With a New Mitral Ring Designed to Simplify Length Determination of Neochords.
PubMed 文献 · Ann Thorac Surg · 2018-02-03

该研究描述一种新型二尖瓣成形环的初步应用经验,该环后部带可拆卸线圈以简化人工腱索长度测定。2015年1月至2016年8月共47例症状性退行性二尖瓣反流患者接受手术,33例(70.2%)经正中胸骨切开、14例(29.8%)经右前外侧小切口。出院时超声显示45例无或仅轻度二尖瓣反流。

要点

  • 新型成形环后部带可拆卸线圈,人工腱索固定于线圈后移除,简化长度测定。
  • 47例退行性二尖瓣反流患者中,中位植入2条人工腱索(最少1条,最多6条)。
  • 出院时45例无或仅轻度二尖瓣反流,平均跨瓣压差2.9±1.3 mm Hg。

核心结论

  • 这种新型成形环便于确定合适的人工腱索长度,并成功用于治疗累及两个瓣叶的不同退行性病变。
  • 该新装置简化了人工腱索的长度测定。

研究设计与样本

  • 单中心初步经验,2015年1月至2016年8月共纳入47例症状性退行性二尖瓣反流患者。
  • 33例(70.2%)经正中胸骨切开,14例(29.8%)经右前外侧小切口植入该半硬质装置。
  • 所有患者术前及体外循环后行经食管超声心动图评估,出院时行经胸超声心动图。

主要发现

  • 中位植入2条人工腱索(最少1条,最多6条)。
  • 联合手术者平均体外循环时间141.7±32.3分钟、主动脉阻断时间104.8±28.5分钟;单纯二尖瓣修复者分别为133±53.9分钟和98.3±41.6分钟。
  • 出院时超声显示45例无或仅轻度二尖瓣反流,平均跨瓣压差2.9±1.3 mm Hg。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Artificial chord implantation has become one of the most applied techniques for mitral valve repair (MVR). Many techniques have been described, with the goal of optimizing neochord implantation. A new annuloplasty device designed to simplify the determination of the appropriate neochord length has been recently introduced. We describe our initial experience with this new device. METHODS: The semirigid device is equipped with removable loops on the posterior aspect of the ring. Neochords are tied to the loops, which are subsequently removed. The device was implanted in 47 symptomatic patients from January 2015 to August 2016 through a median sternotomy in 33 patients (70.2%) and a right anterolateral minithoracotomy in 14 (29.8%). The cause of mitral valve insufficiency was degenerative in all patients, and most patients presented with isolated prolapse of the posterior leaflet. Before and after cardiopulmonary bypass, all patients underwent evaluation with transesophageal echocardiography, and transthoracic echocardiography was performed at discharge. RESULTS: A median of 2 neochords were implanted (minimum, 1; maximum, 6). Mean cardiopulmonary bypass time and aortic cross-clamp times were 141.7 ± 32.3 and 104.8 ± 28.5 minutes for combined and 133 ± 53.9 and 98.3 ± 41.6 minutes for isolated MVR. At discharge, echocardiography revealed no or only mild mitral insufficiency in 45 patients (mean gradient, 2.9 ± 1.3 mm Hg). CONCLUSIONS: This new annuloplasty ring facilitated determination of appropriate neochord length and was used to successfully treat different degenerative pathologies affecting both leaflets. This new device simplified length determination of the neochords.

来源:Ann Thorac Surg,查看原文。版权归原作者所有。

引用

  1. Prinzing A, Bleiziffer S, Krane M, et al. Initial Experience With a New Mitral Ring Designed to Simplify Length Determination of Neochords. Ann Thorac Surg. 2018. PMID: 29408404. doi:10.1016/j.athoracsur.2017.12.030

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。