两种连续流左心室辅助装置流入道位置异常的定量评估

Quantitative Assessment of Inflow Malposition in Two Continuous-Flow Left Ventricular Assist Devices.
PubMed 文献 · Ann Thorac Surg · 2018-01-03

本研究定量评估HeartMate II与HeartWare两种连续流左心室辅助装置流入道套管位置异常及其与装置血栓的关系。分析54例HeartMate II与68例HeartWare患者,发现HeartMate II流入道偏差更大且血栓发生率更高(31%对2.9%,p<0.001)。多变量模型显示HeartMate II及流入道向室间隔偏差增加与更高血栓风险相关(每增加10度比值比1.35)。

要点

  • HeartMate II流入道前平面角度36.7±16.8度,HeartWare为-18.7±11.6度(p<0.001)
  • HeartMate II泵血栓发生率31%,HeartWare为2.9%(p<0.001)
  • 多变量模型中每增加10度向室间隔偏差,血栓风险比值比1.35

核心结论

  • 两种装置的流入道套管位置和血栓发生存在装置依赖性差异,HeartWare位置异常更少且血栓更少。
  • 流入道向室间隔偏移可能通过吸引事件、低流量和转速降低的恶性循环促进泵血栓形成。

研究设计与样本

  • 基于两个正交计算机断层扫描成像平面中与假想理想流入道套管位置的角偏差进行定量评估。
  • 理想位置位于从心尖到二尖瓣中心的连线上。
  • 分析54例HeartMate II患者和68例HeartWare患者。
  • 装置血栓根据临床标准评估。

主要发现

  • HeartMate II流入道前平面角度36.7±16.8度,HeartWare为-18.7±11.6度(p<0.001)。
  • HeartMate II流入道侧平面角度23.7±20.1度,HeartWare为0.2±15.0度(p<0.001)。
  • HeartMate II泵血栓发生率31%,HeartWare为2.9%(p<0.001)。
  • 多变量模型中,HeartMate II和流入道向室间隔偏差增加与更高血栓风险相关(每增加10度比值比1.35)。

关键数值与推荐

  • 前平面正角度表示向游离上壁偏移,负角度表示向下壁偏移。
  • 侧平面正角度表示向室间隔偏移,负角度表示向外侧壁偏移。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: We previously investigated preoperative variables associated with qualitative inflow cannula malposition in the HeartMate II (Thoratec-Abbott, Abbott Park, IL) continuous-flow left ventricular assist device. In this report, we assess inflow cannula malposition quantitatively in recipients of both the HeartMate II and the HeartWare (Medtronic-HeartWare, Minneapolis, MN) and examine its association with device thrombosis. METHODS: Malposition was quantified based on angular deviation from a hypothetic ideal inflow cannula position in two orthogonal computed tomography imaging planes. Ideal position lies on a line from the apex to the center of the mitral valve. Positive anterior plane angulation indicates deviation toward the superior free wall; negative, toward the inferior wall. Positive lateral plane angulation indicates deviation toward the septum; negative, toward the lateral wall. Device thrombosis was assessed based on clinical criteria. RESULTS: Fifty-four HeartMate II patients and 68 HeartWare patients were analyzed. Inflow cannula deviation was significantly higher for HeartMate II than for HeartWare (anterior plane angle 36.7 ± 16.8 versus -18.7 ± 11.6 degrees, p < 0.001; lateral plane angle 23.7 ± 20.1 versus 0.2 ± 15.0 degrees, p < 0.001. Pump thrombosis occurred in 31% of HeartMate II patients and 2.9% of HeartWare patients (p < 0.001). In a multivariate model, HeartMate II and increasing inflow cannula deviation toward the septum were associated with higher thrombosis risk (odds ratio 1.35 per 10-degree increase). CONCLUSIONS: We found distinct device-dependent differences in inflow cannula positioning and thrombosis, with HeartWare showing both less malposition and less thrombosis. Malposition toward the ventricular septum may contribute to pump thrombosis through a vicious cycle of suction events, low flow, and speed reduction.