瓣环尺寸对外科或经导管主动脉瓣置换术后结局的影响
基于CoreValve美国关键性高危试验,按瓣环周长推导直径将患者分为大(≥26 mm)、中(23至<26 mm)、小(<23 mm)瓣环,评估血流动力学、人工瓣膜-患者不匹配(PPM)及临床结局。结果显示小和中瓣环中经导管主动脉瓣置换(TAVR)的平均压差显著低于外科主动脉瓣置换(SAVR),SAVR的PPM发生率与瓣环尺寸显著相关,而TAVR无此关联;瓣环小于26 mm时TAVR血流动力学更优且PPM更少,作者认为选择生物瓣时应强烈考虑TAVR。
要点
- 小和中瓣环中,TAVR术后各次随访平均压差均显著低于SAVR(p < 0.001)。
- SAVR术后PPM发生率与瓣环尺寸显著相关,小瓣环发生率最高;TAVR则无此差异。
- 小和中瓣环中TAVR的PPM显著少于SAVR(p < 0.001),大瓣环两者无差异(p = 0.10)。
核心结论
- 瓣环尺寸对SAVR患者的血流动力学和PPM发生率有显著影响,而TAVR患者未观察到该影响。
- 就瓣环尺寸而言,瓣环小于26 mm时TAVR血流动力学更优且PPM更少,作者认为选择生物瓣时应强烈考虑TAVR。
研究设计与样本
- 基于CoreValve美国关键性高危试验的分析,该试验比较使用自膨胀瓣膜的TAVR与SAVR。
- 采用多层螺旋计算机断层扫描按瓣环周长推导直径将TAVR和SAVR受试者分为大(≥26 mm)、中(23至<26 mm)、小(<23 mm)瓣环,评估血流动力学、PPM和临床结局。
主要发现
- 在小和中瓣环中,TAVR术后各次随访平均压差均显著低于SAVR(p < 0.001)。
- SAVR术后瓣环尺寸与平均压差显著相关,小瓣环压差最高(所有时间点p < 0.05);TAVR术后各瓣环尺寸间压差相似。
- SAVR患者中PPM发生率与瓣环尺寸显著相关,小瓣环发生率最高;TAVR患者中未观察到PPM发生率随瓣环尺寸的差异。
- 小和中瓣环中TAVR患者PPM显著少于SAVR患者(p < 0.001),大瓣环中TAVR与SAVR的PPM发生率无差异(p = 0.10)。
关键数值与推荐
- 瓣环尺寸分类:大瓣环≥26 mm,中瓣环23至<26 mm,小瓣环<23 mm。
- 小和中瓣环中TAVR与SAVR平均压差比较p < 0.001;小和中瓣环中PPM比较p < 0.001;大瓣环中PPM比较p = 0.10。
临床怎么用
- 作者认为,就瓣环尺寸而言,瓣环小于26 mm时TAVR血流动力学更优且PPM更少,选择生物瓣时应强烈考虑TAVR。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: This analysis evaluates the relationship of annular size to hemodynamics and the incidence of prosthesis-patient mismatch (PPM) in surgical aortic valve replacement (SAVR) and transcatheter aortic valve replacement (TAVR) patients. METHODS: The CoreValve US Pivotal High Risk Trial, described previously, compared TAVR using a self-expanding valve with SAVR. Multislice computed tomography was used to categorize TAVR and SAVR subjects according to annular perimeter-derived diameter: large (≥26 mm), medium (23 to <26 mm), and small (<23 mm). Hemodynamics, PPM, and clinical outcomes were assessed. RESULTS: At all postprocedure visits, mean gradients were significantly lower for TAVR compared with SAVR in small and medium size annuli (p < 0.001). Annular size was significantly associated with mean gradient after SAVR, with small annuli having the highest gradients (p < 0.05 at all timepoints); gradients were similar across all annular sizes after TAVR. In subjects receiving SAVR, the frequency of PPM was significantly associated with annular size, with small annuli having the greatest incidence. No difference in PPM incidence by annular sizing was observed with TAVR. In addition, TAVR subjects had significantly less PPM than SAVR subjects in small and medium annuli (p < 0.001), with no difference in the incidence of PPM between TAVR and SAVR in large annuli (p = 0.10). CONCLUSIONS: Annular size has a significant effect on hemodynamics and the incidence of PPM in SAVR subjects, not observed in TAVR subjects. With respect to annular size, TAVR results in better hemodynamics and less PPM for annuli less than 26 mm and should be strongly considered when choosing a tissue valve for small and medium size annuli.