新型转速调节系统用于静脉-动脉体外膜氧合
该研究在六只成年山羊中建立股动静脉VA-ECMO及心源性休克模型,评估一种与自体心动周期同步的转速调节系统。结果显示反搏模式(舒张期增加转速)下左室每搏功和左室收缩末期压力趋于降低,冠状动脉血流趋于升高,提示该单装置可兼具VA-ECMO与IABP效果。
要点
- 反搏模式可能降低VA-ECMO期间左室做功与后负荷。
- 反搏模式可能增加冠状动脉血流。
- 该转速调节系统有望以单一装置整合VA-ECMO与IABP效应。
核心结论
- 该转速调节系统在VA-ECMO期间可实现冠状动脉血流增加及左室做功与后负荷降低。
- 该系统以单一装置提供VA-ECMO与IABP的效果。
研究设计与样本
- 在六只成年山羊中经右房引流和腹主动脉远端灌注建立VA-ECMO。
- 通过输注β-肾上腺素能拮抗剂诱导心源性休克,并在降主动脉放置IABP。
- 在基线、夹闭环路(心源性休克)、连续模式、反搏模式及连续模式联合IABP五种条件下采集数据。
主要发现
- 反搏模式下左室每搏功和左室收缩末期压力趋于降低,提示左室工作负荷与后负荷下降。
- 反搏模式下冠状动脉血流趋于升高。
临床怎么用
- 该新型转速调节系统可能作为股动静脉VA-ECMO期间减轻左室后负荷并改善冠脉血流的单装置方案。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Femoral venoarterial extracorporeal membrane oxygenation (VA-ECMO) is widely used to maintain blood flow in patients with cardiogenic shock. However, retrograde blood flow increases left ventricular (LV) afterload during femoral VA-ECMO. Additional support by means of an intraaortic balloon pump (IABP) alleviates LV afterload but is associated with significant adverse events. We previously developed a system for rotational speed modulation in synchrony with the native cardiac cycle, for use with implantable continuous-flow LV assist devices. Here, we aimed to evaluate whether our novel rotation speed modulation system can improve coronary artery flow and reduce LV during femoral VA-ECMO. METHODS: VA-ECMO was installed by means of right atrial drainage and distal abdominal aortic perfusion in six adult goats. Cardiogenic shock was induced with β-adrenergic antagonist infusion. An IABP was placed in the descending aorta. LV stroke work, LV end-systolic pressure, and coronary arterial flow were evaluated. Data were collected under five conditions (modes): baseline, circuit-clamp (cardiogenic shock), continuous mode (constant rotational speed), counterpulse mode (increasing rotational speed during diastole), and continuous mode with IABP support. RESULTS: LV stroke work and LV end-systolic pressure tended to be lower in the counterpulse mode, indicating decreased LV work load and afterload in this mode. Furthermore, coronary arterial flow tended to be higher in the counterpulse mode. CONCLUSIONS: Our system enabled an increase in coronary arterial flow and a decrease in LV work load and afterload during VA-ECMO. The system offers the effects of VA-ECMO and an IABP in a single device.
引用
- Naito N, Nishimura T, Iizuka K, et al. Novel Rotational Speed Modulation System Used With Venoarterial Extracorporeal Membrane Oxygenation. Ann Thorac Surg. 2017. PMID: 28709666. doi:10.1016/j.athoracsur.2017.04.045