治疗主动脉瓣环扩张的新型装置:离体研究
该离体研究在11个猪主动脉根部单元上测试一种新型装置,由置于主动脉瓣环的内环和主动脉根部水平的外柔性带组成,内环含40%半刚性部分位于心室-动脉交界。在病理扩张条件下,装置使瓣叶对合长度和高度完全恢复至基础值,但跨瓣压差升高、有效瓣口面积下降(P<0.05)。
要点
- 装置由主动脉瓣环内环与根部外柔性带组成,内环40%为半刚性部分。
- 病理扩张下装置使对合长度与高度完全恢复至基础值。
- 处理后压差升高、有效瓣口面积减小,但作者认为无临床相关血流动力学恶化。
核心结论
- 该装置在实验模型中恢复了类生理性对合,未引起临床相关的血流动力学恶化。
研究设计与样本
- 离体研究,使用可容纳猪主动脉瓣的脉动模拟环路,在11个主动脉根部单元上测试。
- 工作条件为心率60 bpm、每搏量75、模拟压力120-80 mmHg。
- 数据在基础、病理(缝合主动脉补片扩张左冠/无冠三角)和装置处理三种条件下采集。
主要发现
- 跨瓣压差与有效瓣口面积:病理条件下分别为0.9 ± 0.64 mmHg和3.1 ± 0.7cm2,处理后分别为3.7 ± 1.1 mmHg和1.5 ± 0.2cm2(P < 0.05)。
- 左冠/无冠直径从病理的2.4 ± 0.2 cm降至处理后的2.0 ± 0.2 cm(P < 0.05)。
- 对合长度和高度在处理后完全恢复至基础值。
- 高速视频和超声心动图确认瓣叶动力学正常。
临床怎么用
- 该装置处于离体实验阶段,尚不能直接用于临床决策。
- 其设计思路为处理主动脉瓣环扩张所致主动脉修复失败提供参考。
局限性
- 研究为离体猪主动脉根部模型,非在体人体研究。
- 样本量为11个主动脉根部单元,结果需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The main reason for aortic repair failures is recurrent annular dilatation. The fibrous portion of left ventricular outflow tract dilates. A novel device was designed to tackle this problem. METHODS: The device consists of an internal ring applied at the aortic annulus plus an external flexible band at the level of the aortic root. The internal ring has a semi-rigid portion (40%, placed at ventriculo-arterial junction) and a flexible portion to allow it to conform along the curves of the non-coronary/right coronary leaflet and right coronary/left coronary leaflet commissures. The external band acts as a reinforcement to the internal ring. A pulsatile mock loop capable of housing porcine aortic valve was used. Working conditions were 60 bpm of heart rate, 75 of stroke volumes and 120-80 mmHg of simulated pressure. Mean gradient, effective orifice area, annular diameter, coaptation height and length were recorded on 11 aortic root units (ARUs). High-speed video and standard echocardiographic images were also recorded. All data were acquired in the following conditions: (i) basal (untreated ARU); (ii) pathological condition (left coronary/non-coronary triangle was dilated by suturing an aortic patch); and (iii) ARU treated with the device. RESULTS: Gradients and effective orifice area were respectively 0.9 ± 0.64 mmHg and 3.1 ± 0.7cm2 (pathological) and 3.7 ± 1.1 mmHg and 1.5 ± 0.2cm2 (treated, P < 0.05). Left coronary/non-coronary diameter decreased from 2.4 ± 0.2 cm (pathological) to 2.0 ± 0.2 (treated, P < 0.05). Coaptation length and height were fully restored to basal values following treatment. Visual inspection showed proper dynamics of the leaflet, confirmed by high-speed video and echocardiography. CONCLUSIONS: The device allowed for restoring physiologic-like coaptation in the experimental model, without inducing clinically relevant worsening of the haemodynamics of the treated ARU.
引用
- Shah P, Romagnoni C, Jaworek M, et al. A novel system for the treatment of aortic annular dilation: an ex vivo investigation. Eur J Cardiothorac Surg. 2017. PMID: 28633398. doi:10.1093/ejcts/ezx203
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。