文献速递心外重症与机械循环支持

Avalon Elite双腔导管混合ECMO保障心脑氧供

Hybrid Extracorporeal Membrane Oxygenation Using Avalon Elite Double Lumen Cannula Ensures Adequate Heart/Brain Oxygen Supply.
PubMed 文献 · Ann Thorac Surg · 2017-05-09

该研究针对外周VA-ECMO合并肺功能受损时的差异性低氧问题,提出在VA-ECMO回路基础上经Avalon Elite双腔导管向肺循环加入VV血流,以提升左心室和升主动脉氧饱和度。研究在7只模拟肺衰竭的成年羊中测试该混合ECMO回路,总ECMO血流2.8至3.3 L/min在VV与VA血流间分配,VV血流从总流量0%按约10%递增至50%。结果显示增加VV血流可逐步提高左心室血氧饱和度,保障心脑氧供。

要点

  • 模拟肺衰竭的VA-ECMO羊出现差异性低氧,左心室氧低而腹主动脉氧高
  • 在VA回路中加入10%至50%的VV血流可逐步提升左心室血氧饱和度
  • VV血流300 mL/min时左心室血氧饱和度82%±6%,1,700 mL/min时达96%±6%

核心结论

  • 基于Avalon Elite双腔导管的混合ECMO回路结构简单、性能高效,且VA/VV血流分配灵活。
  • 递增VV血流(占总ECMO血流10%至50%)可逐步提高左心室血氧水平,确保心脑氧供充足。

研究设计与样本

  • 在7只模拟肺衰竭的成年羊中测试混合ECMO回路。
  • 回路包括27F Avalon Elite双腔导管(下腔静脉-上腔静脉-右心房-下腔静脉)、股动脉17F灌注导管、血泵和气体交换器。
  • 总ECMO血流2.8至3.3 L/min在VV与VA血流间分配,VV血流从0%按约10%递增至总流量50%。

主要发现

  • 纯VA-ECMO伴模拟呼吸衰竭时出现差异性低氧,即左心室氧水平低而腹主动脉氧水平高。
  • 混合ECMO中加入VV血流(占总流量10%至50%)后,左心室血氧饱和度从零VV血流(纯VA-ECMO)时的70%±8%升至300 mL/min VV血流时的82%±6%。
  • 1,700 mL/min VV血流时左心室血氧饱和度达96%±6%。

临床怎么用

  • 对合并肺功能受损的外周VA-ECMO患者,可考虑经Avalon Elite双腔导管加入VV血流以减轻心脑低氧。
  • VV血流可在总ECMO血流的10%至50%范围内按约10%递增调整以滴定左心室氧合。

局限性

  • 研究仅在7只模拟肺衰竭的成年羊中进行,属动物实验,尚未在患者中验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Differential hypoxia exists in peripheral venoarterial (VA) extracorporeal membrane oxygenation (ECMO) patients with compromised lungs, causing hypoxic damage to heart or brain. We proposed an Avalon Elite (Maquet, Rastatt, Germany) double lumen cannula-based hybrid ECMO to add a venovenous (VV) blood flow into the pulmonary circulation onto the existing VA ECMO circuit to increase oxygen saturation in the left ventricle and ascending aorta, mitigating heart/brain hypoxia. METHODS: This hybrid ECMO circuit consists of two cannulas (27F Avalon Elite double lumen cannula from the inferior vena cava to the superior vena cava to right atrium to inferior vena cava; 17F infusion cannula in femoral artery), a blood pump, and a gas exchanger. This hybrid ECMO circuit was tested in 7 adult sheep with simulated lung failure. Total ECMO blood flow (2.8 to 3.3 L/min) was split between VV and VA blood flow. The VV blood flow was adjusted from 0% to 50% of total ECMO flow by approximately 10% increments. RESULTS: In VA ECMO, simulated respiratory failure resulted in differential hypoxia (low oxygen level in left ventricle and high oxygen level in abdominal aorta). In hybrid ECMO, adding VV blood flow (10% to 50% of total ECMO flow) to the VA ECMO circuit progressively increased left ventricle blood oxygen saturation from 70% ± 8% at zero VV blood flow (pure VA ECMO) to 82% ± 6% at 300 mL/min VV blood flow, and 96% ± 6% at 1,700 mL/min VV blood flow. CONCLUSIONS: The Avalon Elite double-lumen cannula-based hybrid ECMO circuit is a simple circuit that provides efficient performance and flexible VA/VV blood distribution. In this hybrid ECMO circuit, incremental VV blood flow (10% to 50%) progressively increased left ventricular blood oxygen level, ensuring adequate heart and brain oxygen supply.