文献速递肺外科

等高与渐变高度吻合钉在支气管闭合中的随机对照研究

Randomized Study Comparing Equal Height Staples With Graduated Height Staples in Bronchial Closure.
PubMed 文献 · Ann Thorac Surg · 2017-05-24

本研究为随机前瞻性研究,纳入61例肺癌肺叶切除术患者,比较等高吻合钉与渐变高度吻合钉在支气管闭合中的成钉效果。结果显示渐变高度组病例评分(2.88 vs 2.17)和完整B形成钉百分比(25.3% vs 10.0%)均显著优于等高组,两组术后并发症无显著差异。

要点

  • 渐变高度吻合钉在支气管残端成钉评分显著优于等高吻合钉。
  • 渐变高度组完整B形成钉比例显著更高(25.3% vs 10.0%)。
  • 两组术后并发症无显著差异,均无支气管胸膜瘘发生。

核心结论

  • 渐变高度吻合钉在支气管残端的成钉评分显著高于等高吻合钉。
  • 从临床角度看,两种吻合钉均表现出可接受的吻合性能。

研究设计与样本

  • 随机前瞻性研究,比较两种吻合钉用于支气管闭合的成钉能力。
  • 共61例患者随机分配至等高吻合钉组(n=30)或渐变高度吻合钉组(n=31)。
  • 评估61例患者的183条吻合线,共1144枚吻合钉。

主要发现

  • 等高吻合钉组病例评分显著低于渐变高度组(2.17 vs 2.88,p=0.0003)。
  • 渐变高度组完整B形成钉百分比显著高于等高组(25.3% vs 10.0%,p=0.000)。
  • 两组术后并发症无显著差异,均未观察到支气管胸膜瘘。

关键数值与推荐

  • 吻合钉成形评分0-4分,4分近似完美的B形钉。

临床怎么用

  • 在肺叶切除术支气管闭合中,渐变高度吻合钉可能提供更优的成钉形态。
  • 两种吻合钉均可接受,临床选择可综合考虑。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: There are two forms of staple cartridge available for lung cancer operations, the flat face with equal height staples and the stepped face with graduated height staples. The objective of the study was to evaluate their stapling ability in lobectomy. METHODS: A randomized prospective study was conducted to compare two types of stapling reloads for bronchial closure. Stapling ability was evaluated by the staple formation on the resected bronchial stump. The stumps were sliced along the staple line, and the shapes of the staples were recorded by roentgenogram. Then, the staple formation was scored as 0 to 4 points with 4 approximating a perfect B-shaped staple. RESULTS: A total of 61 patients were randomly assigned to the equal height staples (n = 30) or the graduated height staples (n = 31). From 61 patients, 183 staple lines, which included 1,144 staples, were evaluated. The case scores were significantly lower in the equal height staples than in the graduated height staples (2.17 versus 2.88, p = 0.0003), respectively. The percentage of staples that formed a complete "B" shape was significantly higher in the graduated height staples than in the equal height staples (25.3% versus 10.0%, p = 0.000). No considerable difference was found between the two groups concerning postoperative complications. No bronchopleural fistula was observed. CONCLUSIONS: The graduated height staples had significantly higher scores of the staple formation on the bronchia stump than the equal height staples. From the clinical point of view, both the equal height staples and the graduated height staples showed acceptable performance in the stapling ability.

来源:Ann Thorac Surg,查看原文。版权归原作者所有。

引用

  1. Okami J, Tokunaga T, Kanou T, et al. Randomized Study Comparing Equal Height Staples With Graduated Height Staples in Bronchial Closure. Ann Thorac Surg. 2017. PMID: 28551048. doi:10.1016/j.athoracsur.2017.02.070

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。