文献速递心外重症与机械循环支持

连续流左心室辅助装置患者腹腔干剪切应力改变的计算与体外分析

Shear stress alterations in the celiac trunk of patients with a continuous-flow left ventricular assist device as shown by in-silico and in-vitro flow analyses.
PubMed 文献 · J Heart Lung Transplant · 2017-03-24

本研究通过计算机流体仿真和3D打印患者特异性体外模型,量化连续流左心室辅助装置(LVAD)对腹腔干及腹主动脉主要分支的血流动力学影响。结果显示左胃动脉平均壁剪切应力(WSS)最高达9.08 ± 5.45 Pa,腹腔干成角与开口远端WSS呈正相关(R = 0.9),可能增加出血风险。

要点

  • 连续流LVAD患者腹腔干分支中左胃动脉WSS最高(9.08 ± 5.45 Pa)。
  • 腹腔干成角与开口远端WSS正相关(R = 0.9),或增加出血风险。
  • 计算流体仿真可辅助临床影像数据评估LVAD患者血流动力学变化。

核心结论

  • 连续流LVAD可改变腹腔干及分支的壁剪切应力,腹腔干成角与开口远端WSS正相关,可能增加血管出血风险。

研究设计与样本

  • 采用计算机流体仿真分析,结合3D打印患者特异性腹腔干模型进行体外流速评估以校准仿真,并计算5例LVAD植入患者的流场与WSS。

主要发现

  • 左胃动脉平均WSS最高,为9.08 ± 5.45 Pa,平均流速0.57 ± 0.25 m/s。
  • 肝总动脉WSS较低,为4.58 ± 1.77 Pa。
  • 腹腔干成角与开口远端WSS呈正相关(R = 0.9)。

临床怎么用

  • 计算流体仿真可补充临床影像数据,用于评估LVAD支持患者的血流动力学改变。

局限性

  • 需更大样本研究确认这些发现。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The use of left ventricular assist devices (LVADs) to treat advanced cardiac heart failure is constantly increasing, although this device leads to high risk for gastrointestinal bleeding. METHODS: Using in-silico flow analysis, we quantified hemodynamic alterations due to continuous-flow LVAD (HeartWare, Inc., Framingham, MA) in the celiac trunk and major branches of the abdominal aorta, and then explored the relationship between wall shear stress (WSS) and celiac trunk orientation. To assess outflow from the aortic branch, a 3-dimensional-printed patient-specific model of the celiac trunk reconstructed from an LVAD-supported patient was used to estimate echocardiographic outflow velocities under continuous-flow conditions, and then to calibrate computational simulations. Moreover, flow pattern and resulting WSS values were computed for 5 patients with LVAD implantation. RESULTS: Peak WSS values were estimated on the 3 branches of the celiac trunk and the LVAD cannula. The mean WSSs demonstrated that the left gastric artery underwent the highest WSS of 9.08 ± 5.45 Pa, with an average flow velocity of 0.57 ± 0.25 m/s compared with that of other vessel districts. The common hepatic artery had a less critical WSS of 4.58 ± 1.77 Pa. A positive correlation was found between the celiac trunk angulation and the WSS stress just distal to the ostium of the celiac trunk (R = 0.9), which may increase vulnerability of this vessel to bleeding. CONCLUSIONS: Although further studies are needed to confirm these findings in a larger patient cohort, computational flow simulations may enhance the information of clinical image data and may have an application in clinical investigations of hemodynamic changes in LVAD-supported patients.

来源:J Heart Lung Transplant,查看原文。版权归原作者所有。

引用

  1. Scardulla F, Pasta S, D'Acquisto L, et al. Shear stress alterations in the celiac trunk of patients with a continuous-flow left ventricular assist device as shown by in-silico and in-vitro flow analyses. J Heart Lung Transplant. 2017. PMID: 28431980. doi:10.1016/j.healun.2017.03.016

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。