瓣中瓣经导管主动脉瓣植入术后主动脉窦血流淤滞的体外研究
该体外研究采用高分辨率粒子图像测速技术,比较外科主动脉瓣与两种经导管主动脉瓣瓣中瓣植入后左冠状动脉窦与非冠状动脉窦的血流动力学变化。结果显示瓣中瓣植入后窦内峰值速度与涡量降低,剪切应力下降,提示血流淤滞可能增加瓣叶血栓风险,非冠状动脉窦更明显。
要点
- 瓣中瓣植入后主动脉窦内流速与涡量降低,提示血流淤滞。
- 非冠状动脉窦瓣叶主动脉侧剪切应力下降更明显,血栓风险可能更高。
- 冠状动脉窦侧经导管瓣叶主动脉侧峰值剪切应力较非冠状动脉窦高约38%。
核心结论
- 瓣中瓣植入后窦内血流与剪切应力降低,提示血流淤滞导致瓣叶血栓风险升高,非冠状动脉窦可能更明显。
研究设计与样本
- 体外研究,以Medtronic Hancock II猪生物瓣模拟外科主动脉瓣,Medtronic CoreValve与Edwards Sapien作为经导管主动脉瓣,采用高分辨率粒子图像测速比较左冠状动脉窦与非冠状动脉窦血流模式。
主要发现
- 外科瓣窦内速度与涡量峰值达0.7 m/s与1000 s-1,非冠状动脉窦瓣叶主动脉侧峰值流体剪切应力下降70%。
- CoreValve瓣中瓣无冠脉血流时峰值速度与涡量约0.4 m/s与550 s-1,有冠脉血流时约0.58 m/s与653 s-1。
- Sapien瓣中瓣无冠脉血流时峰值速度与涡量约0.60 m/s与631 s-1,有冠脉血流时约0.81 m/s与669 s-1。
- 冠状动脉窦侧经导管瓣叶主动脉侧峰值剪切应力较非冠状动脉窦侧高约38%。
临床怎么用
- 瓣中瓣术后应关注血流淤滞相关瓣叶血栓风险,尤其非冠状动脉窦侧。
局限性
- 研究为体外模型,未在体内验证,临床外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Valve-in-valve procedures using transcatheter aortic valves are increasingly performed to treat degenerated bioprosthetic surgical aortic valves because they are less invasive than redo aortic valve replacement. The objective of this study is to quantify the changes in aortic sinus blood flow dynamics before and after a valve-in-valve procedure to gain insight into mechanisms for clinical and subclinical thrombosis of leaflets. METHODS: A detailed description of the sinus hemodynamics for valve-in-valve implantation was performed in vitro. A Medtronic Hancock II (Medtronic Inc, Minneapolis, Minn) porcine bioprosthesis was modeled as a surgical aortic valve, and Medtronic CoreValve and Edwards Sapien (Edwards Lifesciences, Irvine, Calif) valves were used as the transcatheter aortic valves. High-resolution particle image velocimetry was used to compare the flow patterns from these 2 valves within both the left coronary and noncoronary sinuses in vitro. RESULTS: Velocity and vorticity within the surgical valve sinuses reached peak values of 0.7 m/s and 1000 s-1, with a 70% decrease in peak fluid shear stress near the aortic side of the leaflet in the noncoronary sinus. With the introduction of transcatheter aortic valves, peak velocity and vorticity were reduced to approximately 0.4 m/s and 550 s-1 and 0.58 m/s and 653 s-1 without coronary flow and 0.60 m/s and 631 s-1 and 0.81 m/s and 669 s-1 with coronary flow for the CoreValve and Sapien valve-in-valve implantations, respectively. Peak shear stress was approximately 38% higher along the aortic side of the coronary versus noncoronary transcatheter aortic valve leaflet. CONCLUSIONS: Decreased flow and shear stress in valve-in-valve procedures indicate a higher risk of leaflet thrombosis secondary to flow stasis, perhaps more so in the noncoronary sinus.
引用
- Hatoum H, Moore BL, Maureira P, et al. Aortic sinus flow stasis likely in valve-in-valve transcatheter aortic valve implantation. J Thorac Cardiovasc Surg. 2017. PMID: 28433356. doi:10.1016/j.jtcvs.2017.03.053
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。