29毫米第二代球囊扩张式经导管主动脉瓣的支架与瓣叶应力分析

Stent and Leaflet Stresses in 29-mm Second-Generation Balloon-Expandable Transcatheter Aortic Valve.
PubMed 文献 · Ann Thorac Surg · 2017-04-12

本研究通过高分辨率微型CT扫描构建29毫米Edwards Sapien XT经导管主动脉瓣的精确三维几何网格,并在系统压力下用有限元软件分析压缩与未压缩状态的支架和瓣叶应力。结果显示瓣叶峰值应力位于瓣叶与支架连接的连合尖端,该区域最可能发生退变;压缩至26毫米时支架应力显著增高。

要点

  • 瓣叶峰值应力位于连合尖端,是早期退变最可能起始部位。
  • 压缩至26毫米的支架应力远高于未压缩状态,可能影响耐久性。
  • 有限元分析可定位经导管主动脉瓣高应力区域,辅助耐久性评估。

核心结论

  • 29毫米Edwards Sapien XT经导管主动脉瓣的瓣叶峰值应力位于瓣叶附着处的连合尖端,这些区域最可能引发退变。

研究设计与样本

  • 对商用29毫米Edwards Sapien XT瓣膜行高分辨率微型CT扫描构建精确三维几何网格,并用有限元软件在系统压力下模拟压缩与未压缩状态。

主要发现

  • 未压缩瓣膜在舒张压时瓣叶最大和最小主应力分别为1.63 MPa和-0.36 MPa,支架为93.3 MPa和-105.6 MPa。
  • 压缩至26毫米的瓣膜在舒张压时瓣叶最大和最小主应力分别为1.55 MPa和-0.63 MPa,支架为526.1 MPa和-902.2 MPa。
  • 两种状态下瓣叶峰值应力均位于连合尖端,压缩状态下还沿Dacron缝合线出现。

临床怎么用

  • 提示连合尖端是高应力区域,术后随访应关注该区域瓣叶退变迹象。
  • 压缩至较小尺寸可能显著增加支架应力,选择尺寸时需权衡耐久性。

局限性

  • 仅分析两种极端部署几何形态,未涵盖所有植入尺寸和体内动态条件。
  • 材料属性基于钴铬合金和外科生物瓣,可能不完全代表体内实际状态。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Equipoise of transcatheter aortic valve replacement with surgical aortic valve replacement in intermediate-risk patients has been demonstrated. As transcatheter aortic valve replacement usage expands, questions regarding long-term durability become paramount. Valve design impacts durability with regions of increased leaflet stress being vulnerable to early failure. However, transcatheter aortic valve (TAV) leaflet stresses are unknown. The objective of this study was to determine stent and leaflet stresses of second-generation balloon-expandable TAV. METHODS: Commercial 29-mm Edwards Sapien XT (Edwards Lifesciences, Irvine, CA) valves underwent high-resolution microcomputed tomography scanning to develop precise three-dimensional geometric mesh. Compressed and uncompressed TAVs were modeled under systemic pressure using finite element software. Material properties of stent were based on cobalt-chromium, whereas those for leaflets were obtained from surgical bioprostheses. RESULTS: Maximum and minimum principal stresses on uncompressed Sapien XT TAV were 1.63 MPa and -0.36 MPa on leaflets and 93.3 MPa and -105.6 MPa on stent at diastolic pressure. Peak leaflet stress was observed at commissural tips where leaflets connected to the stent. For compressed TAV to 26 mm, maximum and minimum principal stresses were 1.55 MPa and -0.63 MPa on leaflets and 526.1 MPa and -902.2 MPa on stent at diastolic pressure. Peak leaflet stress was located at similar position and also along the suture line with the Dacron (C. R. Bard, Haverhill, PA). CONCLUSIONS: Stress analysis of two extreme deployed geometries of 29-mm Edwards Sapien XT using exact geometry from high-resolution scans demonstrated that peak stresses for TAV leaflets were present at commissural tips where leaflets were attached. These regions would be mostly likely to initiate degeneration.