文献速递心外重症与机械循环支持

左心室辅助装置预后与血流构型及术前左室舒张末内径的交互作用:INTERMACS分析

Left ventricular assist device outcomes based on flow configuration and pre-operative left ventricular dimension: An Interagency Registry for Mechanically Assisted Circulatory Support Analysis.
PubMed 文献 · J Heart Lung Transplant · 2016-12-16

该研究基于INTERMACS注册数据库中9,424例连续流左心室辅助装置(LVAD)植入患者,分析术前左室舒张末内径(LVIDD)与装置血流构型(轴流AC vs 离心CC)的交互作用对不良事件的影响。结果显示该交互项显著预测装置血栓,LVIDD增大时AC装置血栓风险较CC升高,但交互项不预测卒中、消化道出血或生存;CC装置卒中风险更高,而LVIDD每增加1 cm死亡风险降低。

要点

  • 术前LVIDD与血流构型交互项显著预测装置血栓,LVIDD越大AC装置血栓风险较CC越高。
  • CC装置卒中风险较AC更高,与LVIDD无关。
  • LVIDD每增加1 cm死亡风险降低,选择LVAD需综合LVIDD、构型及交互作用。

核心结论

  • 术前LVIDD、血流构型及二者交互项应分别考虑以选择合适LVAD,降低装置血栓、卒中和死亡风险。

研究设计与样本

  • 回顾性注册数据库分析,纳入2008年4月至2015年6月INTERMACS中9,424例连续流LVAD植入患者。
  • 在多变量模型中检验术前LVIDD与血流构型交互项与不良事件的关系。

主要发现

  • 术前LVIDD与血流构型交互项显著预测装置血栓,LVIDD增大时AC装置血栓风险较CC升高(p=0.0099)。
  • LVIDD为7.0 cm时,AC装置血栓风险较CC的HR为1.61(95%CI 1.17-2.22;p=0.004),并随LVIDD增大继续升高。
  • 交互项不预测卒中、消化道出血或患者生存。
  • 多变量模型中,无论LVIDD如何,CC装置卒中风险更高(HR 1.96;95%CI 1.64-2.36;p<0.0001)。
  • 校正分析显示,LVIDD每增加1 cm,死亡风险降低(HR 0.90;95%CI 0.85-0.95;p=0.0004)。

临床怎么用

  • 选择LVAD时需综合术前LVIDD、血流构型及二者交互作用,以降低装置血栓、卒中和死亡风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Axial configuration (AC) and centrifugal configuration (CC) left ventricular assist devices (LVAD) have different flow characteristics, and whether the interaction between device flow configuration and the pre-operative left ventricular internal diastolic diameter (LVIDD) mediates adverse events after LVAD implantation is unknown. METHODS: We queried 9,424 continuous-flow LVAD recipients who received LVADs from April 2008 to June of 2015 in the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS). The pre-operative LVIDD * flow configuration interaction term was tested in multivariable models to determine its relationship to adverse events. RESULTS: The pre-operative LVIDD * flow configuration interaction was a significant predictor of device thrombosis. As the LVIDD increased, the risk of AC device thrombosis increased compared with CC devices (p = 0.0099). At 7.0 cm, the hazard ratio (HR) for AC device thrombosis compared with the CC device was 1.61 (95% confidence interval [CI], 1.17-2.22; p = 0.004) and continued to rise as the LVIDD increased. The LVIDD * flow configuration interaction did not predict stroke, gastrointestinal bleeding, or patient survival. In multivariable models, the hazard of stroke was higher with the CC device regardless of the LVIDD (HR, 1.96; 95% CI, 1.64-2.36; p < 0.0001). Adjusted analyses showed LVAD implantation into a larger left ventricle was associated with a lower risk of death (HR, 0.90; 95% CI, 0.85-0.95; p = 0.0004) per centimeter of LVIDD. CONCLUSIONS: Our study suggests that the pre-operative LVIDD, flow configuration, and interaction terms should be considered individually when choosing the appropriate LVAD to mitigate the rates of device thrombosis, stroke, and death.

来源:J Heart Lung Transplant,查看原文。版权归原作者所有。

引用

  1. Shah P, Birk S, Maltais S, et al. Left ventricular assist device outcomes based on flow configuration and pre-operative left ventricular dimension: An Interagency Registry for Mechanically Assisted Circulatory Support Analysis. J Heart Lung Transplant. 2016. PMID: 28087105. doi:10.1016/j.healun.2016.12.004