胸主动脉腔内修复对离体猪主动脉径向应变的影响

Impact of thoracic endovascular aortic repair on radial strain in an ex vivo porcine model.
PubMed 文献 · Eur J Cardiothorac Surg · 2017-01-01

研究利用20个新鲜猪胸主动脉在模拟循环回路中量化胸主动脉腔内修复(TEVAR)对径向主动脉应变的影响。结果显示TEVAR使支架段径向应变降低49.4±24.0%(P<0.001),导致支架段与近端非支架段应变不匹配(7.0±2.5% vs 11.8±3.9%,P<0.001),而远端非支架段未受影响(P=0.99)。支架移植物oversizing未显著影响应变降低程度(P=0.30)。

要点

  • TEVAR使胸主动脉僵硬度增加2倍,支架段径向应变显著降低。
  • 支架段与近端非支架段出现应变不匹配,可能影响Windkessel功能。
  • 支架移植物oversizing程度不影响应变降低量。

核心结论

  • TEVAR使胸主动脉僵硬度增加2倍,节段性僵化可能显著削弱Windkessel功能,并与支架诱导的夹层和动脉瘤扩张等并发症相关。

研究设计与样本

  • 离体实验研究,使用20个新鲜猪胸主动脉连接至离心泵驱动的模拟循环回路。
  • 在TEVAR前后于5个压力水平(100、120、140、160、180 mmHg)拍摄直径,按oversizing分为0-9%(n=7)、10-19%(n=6)、20-29%(n=6)三组。
  • 径向应变由直径振幅除以100 mmHg下的基线直径得出,并通过单轴拉伸测试评估杨氏模量。

主要发现

  • TEVAR后支架段内径向应变降低49.4±24.0%(P<0.001)。
  • 支架段与近端非支架段之间出现应变不匹配(7.0±2.5% vs 11.8±3.9%,P<0.001),而远端非支架段未受影响(P=0.99)。
  • 支架移植物oversizing未显著影响应变降低量(P=0.30)。
  • 拉伸测试显示胸主动脉近端趋向于比远端更具弹性(P=0.11)。

临床怎么用

  • TEVAR可导致胸主动脉节段性僵化,临床需关注支架段与近端非支架段的应变不匹配。
  • 该僵化可能削弱Windkessel功能,并与支架诱导的夹层和动脉瘤扩张等并发症相关。
  • 研究数据可能对未来的支架移植物设计有启示,尤其是针对高顺应性的近端胸主动脉TEVAR。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To quantify the impact of thoracic endovascular aortic repair (TEVAR) on radial aortic strain with the aim of elucidating stent-graft-induced stiffening and complications. METHODS: Twenty fresh thoracic porcine aortas were connected to a mock circulatory loop driven by a centrifugal flow pump. A high-definition camera captured diameters at five different pressure levels (100, 120, 140, 160, and 180 mmHg), before and after TEVAR. Three oversizing groups were created: 0-9% ( n = 7), 10-19% ( n = 6), and 20-29% ( n = 6). Radial strain (or deformation) derived from diameter amplitude divided by baseline diameter at 100 mmHg. Uniaxial tensile testing evaluated Young's moduli of the specimens. RESULTS: Radial strain was reduced after TEVAR within the stented segment by 49.4 ± 24.0% ( P < 0.001). As result, a strain mismatch was observed between the stented segment and the proximal non-stented segment (7.0 ± 2.5% vs 11.8 ± 3.9%, P < 0.001), whereas the distal non-stented segment was unaffected ( P = 0.99). Stent-graft oversizing did not significantly affect the amount of strain reduction ( P = 0.30). Tensile testing showed that the thoracic aortas tended to be more elastic proximally than distally ( P = 0.11). CONCLUSIONS: TEVAR stiffened the thoracic aorta by 2-fold. Such segmental stiffening may diminish the Windkessel function considerably and might be associated with TEVAR-related complications, including stent-graft-induced dissection and aneurysmal dilatation. These data may have implications for future stent-graft design, in particular for TEVAR of the highly compliant proximal thoracic aorta.