文献速递主动脉外科

应用新型斑点追踪超声心动图获取升主动脉瘤生物力学行为

Obtaining the biomechanical behavior of ascending aortic aneurysm via the use of novel speckle tracking echocardiography.
PubMed 文献 · J Thorac Cardiovasc Surg · 2016-12-20

研究纳入17例接受升主动脉切除患者,平均主动脉直径56.16±15 mm,术中采用斑点追踪经食管超声心动图(TEE)计算体内硬度指数,并与离体组织力学测试结果比较。结果显示体内硬度指数与离体25%表观硬度在四个象限均无显著差异(P=.8617),提示斑点TEE有望用于估测升主动脉组织的离体力学特性。

要点

  • 斑点追踪TEE可在术中计算升主动脉体内硬度指数。
  • 体内硬度指数与离体组织力学测试结果无显著差异。
  • 该方法有望辅助评估扩张升主动脉的破裂或夹层风险。

核心结论

  • 斑点追踪TEE似乎是估测升主动脉组织离体力学特性的一种有前景的技术。

研究设计与样本

  • 前瞻性招募17例接受升主动脉切除的患者,平均主动脉直径56.16±15 mm。
  • 术中行斑点追踪TEE计算体内硬度指数,并与离体组织力学测试的25%表观硬度比较。

主要发现

  • 体内硬度指数在内曲度、前壁、外曲度和后壁分别为0.0544±0.0490、0.0295±0.0199、0.0411±0.0328和0.0502±0.0320。
  • 离体25%表观硬度在内曲度、前壁、外曲度和后壁分别为0.0616±0.0758 MPa、0.0352±0.00992 MPa、0.0405±0.0199 MPa和0.0327±0.0106 MPa。
  • 患者匹配的离体25%表观硬度与体内硬度指数无显著差异(P=.8617,重复测量双因素方差分析)。

临床怎么用

  • 斑点追踪TEE或可用于术中无创评估升主动脉硬度,辅助判断扩张主动脉的破裂或夹层风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Ex vivo measurement of ascending aortic biomechanical properties may help understand the risk for rupture or dissection of dilated ascending aortas. A validated in vivo method that can predict aortic biomechanics does not exist. Speckle tracking transesophageal echocardiography (TEE) has been used to measure ventricular stiffness; we sought to determine whether speckle TEE could be adapted to estimate aortic stiffness in vivo and compare these findings with those obtained by ex vivo tissue measurements. METHODS: A total of 17 patients undergoing ascending aortic resection were recruited to with a mean aortic diameter was 56.16 ± 15 mm. Intraoperative speckle TEE tracking analysis was used to calculate aortic stiffness index using the following equation: β2=ln(SBP/DBP)/AoS, where β2 is the stiffness index; SBP is systolic blood pressure; DBP is diastolic blood pressure; and AoS is the circumferential strain. Ex vivo stiffness was obtained by mechanical tissue testing according to previously described methods. The aortic ring at the pulmonary trunk was divided into 4 equal quadrants. RESULTS: The in vivo stiffness index for the inner curvature, anterior wall, outer curvature, and posterior wall were 0.0544 ± 0.0490, 0.0295 ± 0.0199, 0.0411 ± 0.0328, and 0.0502 ± 0.0320, respectively. The mean ex vivo 25% apparent stiffness for inner curvature, anterior wall, outer curvature, and posterior wall were 0.0616 ± 0.0758 MPa, 0.0352 ± 0.00992 MPa, 0.0405 ± 0.0199 MPa, and 0.0327 ± 0.0106 MPa, respectively. The patient-matched ex vivo 25% apparent stiffness and in vivo stiffness index were not significantly different (P = .8617, 2-way analysis of variance with repeated measures). CONCLUSIONS: The use of speckle TEE appears to be a promising technique to estimate ex vivo mechanical properties of the ascending aortic tissue.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Alreshidan M, Shahmansouri N, Chung J, et al. Obtaining the biomechanical behavior of ascending aortic aneurysm via the use of novel speckle tracking echocardiography. J Thorac Cardiovasc Surg. 2016. PMID: 28094007. doi:10.1016/j.jtcvs.2016.11.056

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。