26毫米首代球囊扩张式经导管主动脉瓣的支架与瓣叶应力分析

Stent and leaflet stresses in a 26-mm first-generation balloon-expandable transcatheter aortic valve.
PubMed 文献 · J Thorac Cardiovasc Surg · 2016-12-23

本研究通过高分辨率微计算机断层扫描获取26毫米Sapien经导管心脏瓣膜的精确三维几何,并用有限元分析在80和120毫米汞柱压力下计算支架与瓣叶应力。结果显示瓣叶最大主应力为1.31兆帕,支架最大主应力为188.91兆帕,峰值均位于瓣叶与支架连接的连合处;非圆柱形几何使瓣叶峰值应力增加16%。这些区域可能易发生瓣叶退变。

要点

  • 26毫米Sapien瓣叶在80毫米汞柱下最大主应力为1.31兆帕。
  • 支架最大主应力188.91兆帕,瓣叶与支架峰值应力均位于连合处。
  • 非圆柱形几何使瓣叶峰值主应力增加16%。

核心结论

  • 26毫米Sapien经导管心脏瓣膜的支架和瓣叶峰值应力均位于瓣叶附着的连合处,这些区域可能易发生瓣叶退变。

研究设计与样本

  • 对26毫米Sapien经导管心脏瓣膜进行高分辨率微计算机断层扫描,构建瓣叶、支架和聚对苯二甲酸乙二醇酯元件的精确三维几何,并采用有限元分析在80和120毫米汞柱压力下计算应力。

主要发现

  • 在80毫米汞柱下,Sapien瓣叶最大主应力为1.31兆帕,峰值位于瓣叶与支架连接的连合尖端。
  • 支架最大主应力为188.91兆帕,位于瓣叶连合附着的支架尖端。
  • 非圆柱形几何使瓣叶峰值主应力增加16%。

临床怎么用

  • 了解首代经导管瓣膜的应力分布有助于与未来设计比较相对耐久性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Transcatheter aortic valve replacement is established therapy for high-risk and inoperable patients with severe aortic stenosis, but questions remain regarding long-term durability. Valve design influences durability. Increased leaflet stresses in surgical bioprostheses have been correlated with degeneration; however, transcatheter valve leaflet stresses are unknown. From 2007 to 2014, a majority of US patients received first-generation balloon-expandable transcatheter valves. Our goal was to determine stent and leaflet stresses in this valve design using finite element analyses. METHODS: A 26-mm Sapien Transcatheter Heart Valve (Edwards Lifesciences, Inc, Irvine, Calif) underwent high-resolution microcomputed tomography scanning to develop precise 3-dimensional geometry of the leaflets, the stent, and the polyethylene terephthalate elements. The stent was modeled using 3-dimensional elements and the leaflets were modeled using shell elements. Stent material properties were based on stainless steel, whereas those for leaflets were obtained from surgical bioprostheses. Noncylindrical Sapien valve geometry was also simulated. Pressure loading to 80 mm Hg and 120 mm Hg was performed using ABAQUS finite element software (Dassault Systèmes, Waltham, Mass). RESULTS: At 80 mm Hg, maximum principal stresses on Sapien leaflets were 1.31 megaspascals (MPa). Peak leaflet stress was observed at commissural tips where leaflets connected to the stent. Maximum principal stresses for the stent were 188.91 MPa and located at stent tips where leaflet commissures were attached. Noncylindrical geometry increased peak principal leaflet stresses by 16%. CONCLUSIONS: Using exact geometry from high-resolution scans, the 26-mm Sapien Transcatheter Heart Valve showed that peak stresses for both stent and leaflets were present at commissural tips where leaflets were attached. These regions would be prone to leaflet degeneration. Understanding stresses in first-generation transcatheter valves allows comparison to future designs for relative durability.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Xuan Y, Krishnan K, Ye J, et al. Stent and leaflet stresses in a 26-mm first-generation balloon-expandable transcatheter aortic valve. J Thorac Cardiovasc Surg. 2016. PMID: 28108064. doi:10.1016/j.jtcvs.2016.12.016

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。