文献速递心外重症与机械循环支持

脓毒症猪血在体外膜氧合中未进一步激活凝血

Septic porcine blood does not further activate coagulation during in vitro membrane oxygenation.
PubMed 文献 · Eur J Cardiothorac Surg · 2017-01-01

该研究利用体外12小时ECMO模拟环路,对比脓毒症猪血与健康猪血的凝血和炎症参数。脓毒症血初始即呈高凝和炎症状态,但接触膜氧合器后凝血活性、炎症反应及氧合器阻力均未进一步增加。提示预激活的脓毒症血与氧合器接触不额外加重凝血或生物相容性问题。

要点

  • 脓毒症猪血初始凝血和炎症指标显著高于健康血。
  • 接触膜氧合器后脓毒症血凝血活性未进一步升高。
  • 未观察到氧合器阻力增加或血栓形成增多。

核心结论

  • 脓毒症猪血接触氧合器后未进一步激活凝血,未见血栓形成或生物相容性问题增加。

研究设计与样本

  • 采用先前建立的体外12小时ECMO模拟环路,使用脓毒症或健康家猪血分析凝血和炎症参数。
  • 脓毒症通过盲肠结扎穿孔模型在猪中诱导。

主要发现

  • 实验开始时脓毒症血凝血酶-抗凝血酶复合物为76.9 µg/l,健康血为27.7 µg/l。
  • 脓毒症血D-二聚体为1.2 mg/l,健康血为0.3 mg/l。
  • 脓毒症血纤维蛋白原浓度为1.8 g/l,健康血为1.5 g/l。
  • 脓毒症血EXTEM-CT为44.2 s,健康血为57 s。
  • 脓毒症血乳酸为3.4 mmol/l,健康血为1.5 mmol/l;白介素-6为827 pg/ml,健康血为31 pg/ml。

临床怎么用

  • 脓毒症患者行ECMO时,预激活的凝血和炎症状态可能不因氧合器接触而进一步恶化。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: For patients with a severe acute respiratory distress syndrome (ARDS), extracorporeal membrane oxygenation (ECMO) represents a life-saving measure. Frequently, patients with severe ARDS also show signs of severe sepsis. As blood contact with the membrane oxygenator surface leads to adverse effects due to insufficient biocompatibility partly caused by activation of platelets, coagulation factors and leucocytes, we hypothesized that these adverse effects would be amplified if septic blood in a preactivated state came into contact with the membrane oxygenator. METHODS: In a previously established in vitro 12-h ECMO test system (mock loop), we used septic or healthy domestic pig blood to analyse coagulation and inflammatory parameters. Sepsis was induced by a caecal ligation and puncture model in pigs. RESULTS: At the beginning of the mock loop experiments, the septic blood showed significantly increased thrombin-antithrombin complexes (76.9 vs 27.7 µg/l), D-dimers (1.2 vs 0.3 mg/l) and fibrinogen concentration (1.8 vs 1.5 g/l), as well as elevated extrinsic coagulation activity (shorter EXTEM-CT: 44.2 vs 57 s) and higher lactate (3.4 vs 1.5 mmol/l) and cytokine levels (interleukin-6: 827 vs 31 pg/ml) when compared with the blood from healthy animals. Despite the preactivated status of the septic blood, no further increase of coagulation activity, inflammatory response or increased oxygenator resistance was observed in comparison to the control experiments. CONCLUSION: Septic porcine blood was not further activated due to the contact with an oxygenator, and no increased clot formation or biocompatibility problems were observed.

来源:Eur J Cardiothorac Surg,查看原文。版权归原作者所有。

引用

  1. Bleilevens C, Grottke O, Groening S, et al. Septic porcine blood does not further activate coagulation during in vitro membrane oxygenation. Eur J Cardiothorac Surg. 2017. PMID: 27806995. doi:10.1093/ejcts/ezw345