非动力吻合器在重度肺气肿肺减容术中的疗效:前瞻性随机单盲单中心研究
本研究纳入19例患者共32次肺减容手术,随机使用非动力AEON Endostapler或动力Echelon Flex Powered Plus Stapler,通过数字记录系统监测术后漏气。非动力组即刻漏气发生率6/17(35.3%),动力组9/15(60%);漏气闭合中位时间非动力组14.3小时[6.7, 116],动力组93.2小时[2.1, 159];两种吻合器均可行,非动力吻合器显示更早漏气闭合和更低漏气发生率。
要点
- 非动力吻合器组术后即刻漏气发生率35.3%,低于动力组的60%。
- 非动力组漏气闭合中位时间14.3小时,短于动力组的93.2小时。
- 两种吻合器在重度肺气肿肺减容术中均可行,但差异未达统计学显著。
核心结论
- 两种吻合器系统均可用于重度肺气肿患者的肺减容手术。
- 非动力吻合器显示更早的漏气闭合和更低的术后漏气发生率。
研究设计与样本
- 前瞻性随机单盲单中心研究,共32次肺减容手术,来自19例患者,按手术侧分层随机分配至非动力或动力吻合器组。
- 术后漏气通过数字记录系统监测,评估漏气闭合时间、漏气发生率和严重程度及胸管留置时间。
主要发现
- 非动力吻合器组即刻术后漏气6/17(35.3%),动力吻合器组9/15(60%)。
- 漏气闭合中位时间非动力组14.3小时[6.7, 116],动力组93.2小时[2.1, 159]。
临床怎么用
- 在重度肺气肿肺减容术中可考虑使用非动力吻合器以减少术后漏气并缩短漏气时间。
- 需结合患者具体情况和手术条件选择吻合器类型。
局限性
- 样本量小(19例患者32次手术),单中心设计。
- 单盲设计可能引入偏倚。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This study aimed to assess whether the use of the non-powered AEON Endostapler during lung resection in patients with severe lung emphysema reduces the duration of postoperative air leak, as measured by the air leak volume over time, in comparison to the Echelon Flex Powered Plus Stapler. METHODS: A total of 32 lung volume reduction surgeries were performed on 19 patients, stratified by side of the operation. These procedures were randomly assigned to utilize either the non-powered or the powered stapler. Postoperative air leak was monitored using a digital recording system. The time to air leak closure, the incidence and severity of air leaks, and the duration of chest tube placement were evaluated. RESULTS: Immediate postoperative air leaks were observed in 6 of 17 procedures (35.3%) using the non-powered stapler and in 9 of 15 procedures (60%) performed with the powered stapler. The median time to closure of the air leak was also considerably shorter in the non-powered procedures: 14.3 hours [6.7, 116] compared to 93.2 hours [2.1, 159] for the powered treatments. Cox regression analysis yielded a hazard ratio of 1.6 (95% CI, 0.73-3.3) for a faster air leak closure with the non-powered stapler (P = .25). CONCLUSIONS: Both stapler systems are feasible for use in patients with severe lung emphysema. However, our results show an earlier air leak closure and a lower incidence of postoperative air leaks with the use of a non-powered stapler. CLINICAL REGISTRATION NUMBER: https://clinicaltrials.gov/study/NCT05628415? term=LVRS&rank=9, 28.11.2022.
引用
- Macharia-Nimietz EF, Mallaev M, Gecas P, et al. Efficacy of Non-Powered Stapler in Lung Volume Reduction Surgery of Severe Lung Emphysema: A Prospective Randomized Single-Blinded Monocentric Study. Interdiscip Cardiovasc Thorac Surg. 2026. PMID: 41787969. doi:10.1093/icvts/ivag071
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。