内几何瓣环成形环主动脉瓣修复的中期结局
结论认为该环虽能有效减小瓣环尺寸,但刚性环内设计及过度瓣环缩小可能易致复发性关闭不全并限制未来经导管选项。
要点
- 内几何瓣环成形环修复主动脉瓣后44%患者需再干预,5年再干预率40.8%。
- 再干预主要因严重主动脉瓣关闭不全,多数行外科主动脉瓣置换。
- 刚性环内设计及过度瓣环缩小可能增加复发性关闭不全风险。
核心结论
- 使用内几何瓣环成形环行主动脉瓣修复后44%患者需再干预,5年估计再干预率40.8%(95%CI 20.0%-70.5%)。
- 该环虽有效减小瓣环尺寸,但刚性环内设计及过度瓣环缩小可能易致复发性关闭不全并限制未来经导管选项。
研究设计与样本
- 回顾性 review,纳入2017年1月至2025年5月所有使用内几何瓣环成形环行主动脉瓣修复的成年患者。
- 平均随访3.7±1.7年。
主要发现
- 7例患者(44%)需8次再干预,原因为严重主动脉瓣关闭不全(n=7)或感染性心内膜炎伴中度关闭不全(n=1)。
- 再干预方式包括外科主动脉瓣置换(n=5)、再次内几何瓣环成形(n=1)、主动脉根部置换(n=1)及瓣中环经导管主动脉瓣置换(n=1)。
- 外科病例中,取出的环的瓣环尺寸一致恢复至植入前大小。
临床怎么用
- 使用该环行主动脉瓣修复后需密切随访,警惕复发性主动脉瓣关闭不全及再干预风险。
- 考虑该环刚性环内设计及过度瓣环缩小可能限制未来经导管选项,应谨慎评估其广泛应用。
局限性
- 原文未报告其他局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The internal geometric annuloplasty ring is the only commercially available aortic annuloplasty ring designed to reduce and prevent annular dilatation during aortic valve repair; however, data regarding its long-term durability remain limited. Here we present our mid-term outcomes. METHODS: A retrospective review was conducted of all adult patients who underwent aortic valve repair using the internal geometric annuloplasty ring between January 2017 and May 2025. Patients requiring valve reintervention were included in the analysis. RESULTS: Eighteen aortic valve repairs were performed in 17 patients (mean age, 54 ± 15 years; 16 males; 14 tricuspid valves, 2 bicuspid valves, 1 unicuspid valve), including 1 redo repair. Ten cases (56%) involved concomitant aortic root remodeling. Tricuspid 21-mm rings were used most frequently (61%). During a mean follow-up of 3.7 ± 1.7 years, 8 reinterventions were required in 7 patients (44%) owing to severe aortic insufficiency (AI; n = 7) or endocarditis with moderate AI (n = 1). The 5-year estimated reintervention rate was 40.8% (95% confidence interval, 20.0%-70.5%). Reinterventions included surgical aortic valve replacement (n = 5), redo internal geometric annuloplasty (n = 1), aortic root replacement (n = 1), and valve-in-ring transcatheter aortic valve replacement (n = 1). In surgical cases, postexplantation annular dimensions of the ring consistently returned to their preimplantation sizes. CONCLUSIONS: In our experience, 44% of patients receiving the internal geometric annuloplasty ring required aortic valve reintervention. Although effective in reducing annular size, the ring's rigid intra-annular design and the potential for excessive annular downsizing may predispose to recurrent insufficiency and limit future transcatheter options, prompting careful consideration of its broad application.
引用
- Yokoyama Y, Hawkins RB, Hamilton BCS, et al. Mid-term outcomes after aortic valve repair with the internal geometric annuloplasty ring. J Thorac Cardiovasc Surg. 2026. PMID: 41864277. doi:10.1016/j.jtcvs.2026.03.573
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。