文献速递主动脉外科

双分支覆膜支架四阶段释放全弓修复猪模型安全性与可行性

Double-branched stent graft and four-stage deployment in total arch repair: safety and feasibility evaluation in porcine models.
PubMed 文献 · Interdiscip Cardiovasc Thorac Surg · 2024-01-01

本研究在12头健康猪中评估新型双分支覆膜支架系统采用四阶段释放技术行主动脉弓修复的安全性与可行性,每组6头,分别于术后90天和180天取样。技术成功率100%(12/12),全部动物存活至实验终点,随访影像未见内漏或器械移位,病理证实生物相容性满意。

要点

  • 双分支覆膜支架四阶段释放技术在全弓修复猪模型中技术成功率100%(12/12)。
  • 全部动物存活至实验终点,术后随访未见主要不良心血管事件。
  • 随访终点影像未见内漏及器械移位,病理证实生物相容性满意。

核心结论

  • 依据猪模型的临床前评估,该新型双分支覆膜支架系统联合四阶段释放技术被认定为主动脉弓修复的安全且可行方案。

研究设计与样本

  • 研究在12头健康猪中评估双分支覆膜支架系统(含近端涤纶人工血管、主体及双分支覆膜支架和输送系统)行主动脉弓修复,每组6头。
  • 术后分别于90天和180天取样,行术前术后影像学及术中动脉血气分析,处死后取植入物、周围组织及主要器官行病理分析。

主要发现

  • 支架植入技术成功率为100%(12/12),全部动物存活至实验终点。
  • 围术期评估显示支架完整,随访终点影像未见内漏及器械移位。
  • 术后随访期间未观察到主要不良心血管事件,病理检查证实支架生物相容性满意。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The primary objective of this research was to evaluate the safety and feasibility of an innovative double-branched stent graft system employing four-stage deployment technology for aortic arch repair in porcine models. METHODS: The double-branched stent graft system consisted of a proximal polyester artificial blood vessel, the main and double-branched stent grafts and a delivery system. We utilized 12 healthy pigs as experimental animals (6 per group). Postimplantation, samples were collected at 90 and 180 days after the operations. Preoperative and postoperative imaging and intraoperative arterial blood gas analyses were performed. After the pigs were euthanized, the implanted product, surrounding tissue and major organs were collected for pathological analysis. RESULTS: The technical success rate of the stent graft implants was 100% (12/12). All animals survived to the experimental end point. Perioperative assessments showed intact stent grafts, and imaging features at the end of the follow-up period revealed neither endoleak nor device migration. No major adverse cardiovascular events were observed during the postoperative follow-up period. Pathological examinations confirmed the satisfactory biocompatibility of the stent graft. CONCLUSIONS: This innovative double-branched stent graft system with four-stage deployment technology was affirmed as a safe and feasible option for aortic arch repair in accordance with our preclinical evaluation with porcine models.

来源:Interdiscip Cardiovasc Thorac Surg,查看原文。版权归原作者所有。

引用

  1. Wang C, Li W, Yang P, et al. Double-branched stent graft and four-stage deployment in total arch repair: safety and feasibility evaluation in porcine models. Interdiscip Cardiovasc Thorac Surg. 2024. PMID: 38492560. doi:10.1093/icvts/ivae049

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。