文献速递瓣膜外科

术前CT规划用于外科主动脉瓣置换的瓣膜尺寸选择

Preprocedural Computed Tomography Planning for Surgical Aortic Valve Replacement.
PubMed 文献 · Ann Thorac Surg · 2024-02-19

本研究比较了CT瓣环测量指导与常规术中测瓣器选择外科主动脉瓣置换瓣膜尺寸的效果,CT组植入尺寸小于CT预测尺寸的比例和重度患者-人工瓣膜不匹配率均显著低于常规组。CT sizing有望在术前预测不匹配并识别需行主动脉根部加宽的患者。

要点

  • CT sizing组植入尺寸小于CT预测尺寸的比例为12%,显著低于常规组的31%(P=.001)。
  • CT sizing组重度患者-人工瓣膜不匹配率为0%,显著低于常规组的6%(P=.039)。
  • 常规 sizing下操作者间变异与植入尺寸偏小相关,CT sizing后该关联不再显著。

核心结论

  • 将CT sizing应用于外科主动脉瓣置换可改善瓣膜尺寸选择、减少患者-人工瓣膜不匹配并降低操作者间变异。
  • CT sizing还可用于术前预测患者-人工瓣膜不匹配并识别需要主动脉根部加宽的患者。

研究设计与样本

  • 本研究纳入使用Inspiris或Magna Ease行外科主动脉瓣置换且术前接受心电图门控CT成像的患者。
  • 自2022年6月起,研究者应用基于CT瓣环尺寸的CT sizing算法指导最小外科瓣膜标签尺寸,最终瓣膜选择由术中外科医生决定。
  • 研究比较了两个队列的瓣膜选择适当性(植入尺寸与CT预测尺寸比较)及未行主动脉根部加宽情况下的患者-人工瓣膜不匹配率:2022年6月以来的102例(CT sizing队列)与2020至2021年的180例(常规 sizing队列)。

主要发现

  • 植入尺寸小于CT预测尺寸的比例:CT sizing组12% vs 常规 sizing组31%(P=.001),CT sizing组显著更低。
  • 重度患者-人工瓣膜不匹配率:CT sizing组0% vs 常规 sizing组6%(P=.039),CT sizing组显著更低。
  • 常规 sizing下操作者间变异是植入尺寸小于CT预测尺寸的相关因素,而CT sizing后该关联不再显著。

关键数值与推荐

  • 植入尺寸小于CT预测尺寸:CT sizing队列12% vs 常规 sizing队列31%(P=.001)。
  • 重度患者-人工瓣膜不匹配:CT sizing队列0% vs 常规 sizing队列6%(P=.039)。

临床怎么用

  • 作者结论认为,将CT sizing应用于外科主动脉瓣置换可改善瓣膜尺寸选择、减少患者-人工瓣膜不匹配并降低操作者间变异。
  • 作者结论认为,CT sizing可用于术前预测患者-人工瓣膜不匹配并识别需要主动脉根部加宽的患者。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Selection of transcatheter valve size using preprocedural computed tomography (CT) is standardized and well established. However, valve sizing for surgical aortic valve replacement (SAVR) is currently performed intraoperatively by using sizers, which may result in variation among operators and risk for prosthesis-patient mismatch. This study evaluated the usefulness of CT annulus measurement for SAVR valve sizing. METHODS: This study included patients who underwent SAVR using Inspiris or Magna Ease and received preoperative electrocardiogram-gated CT imaging. Starting from June 2022, study investigators applied a CT sizing algorithm using CT-derived annulus size to guide minimum SAVR label size. The final decision of valve selection was left to the operating surgeon during SAVR. The study compared the appropriateness of valve selection (comparing implanted size with CT-predicted size) and prosthesis-patient mismatch rates without aortic root enlargement between 2 cohorts: 102 cases since June 2022 (CT sizing cohort) and 180 cases from 2020 to 2021 (conventional sizing cohort). RESULTS: Implanted size smaller than CT predicted size and severe prosthesis-patient mismatch were significantly lower by CT sizing than by conventional sizing (12% vs 31% [P = .001] and 0% vs 6% [P = .039], respectively). Interoperator variability was a factor associated with implanted size smaller than CT predicted with conventional sizing, whereas it became nonsignificant with CT sizing. CONCLUSIONS: Applying CT sizing to SAVR led to improved valve size selection, less prosthesis-patient mismatch, and less interoperator variability. CT sizing for SAVR could also be used to predict prosthesis-patient mismatch before SAVR and identify patients who need aortic root enlargement.