可修复的功能性与退行性二尖瓣反流体外模型
研究利用猪二尖瓣和搏动流循环系统建立功能性与退行性二尖瓣反流体外模型。功能性模型经0.1%胶原酶浸泡及3D打印扩张器扩张,退行性模型通过切断P2腱索并调整切断数量控制反流程度;两种模型均可通过缘对缘或人工腱索技术修复,修复后反流分数显著降低。
要点
- 功能性模型中重度反流分数为47.9%(SD 2.2%),修复后降至28.7%(SD 4.4%)(P < 0.001)。
- 退行性模型中重度反流分数为58.5%(SD 8.0%),修复后降至26.0%(SD 4.4%)(P < 0.001)。
- 功能性重度模型平均反流分数为59.4%(SD 6.0%),两种模型均可产生满足介入适应症标准的反流水平。
核心结论
- 功能性与退行性模型均可产生满足介入适应症标准的充分二尖瓣反流水平。
- 可修复模型对评估瓣膜修复手术与器械疗效有价值,反流程度可控提升模型通用性与可靠性。
- 这些可重复模型可加速新型器械研发。
研究设计与样本
- 研究使用猪二尖瓣与搏动流循环系统建立两种可修复体外模型。
- 功能性模型将瓣环浸于0.1%胶原酶溶液并用3D打印扩张器扩张,通过调整扩张器及缝合瓣膜的硅胶片尺寸控制反流分级。
- 退行性模型切断P2腱索,通过调整切断腱索数量控制反流分级。
- 模型采用缘对缘或人工腱索技术修复。
主要发现
- 功能性模型中重度模型平均反流分数为47.9%(SD 2.2%),修复后降至28.7%(SD 4.4%)(P < 0.001)。
- 退行性模型中重度模型平均反流分数为58.5%(SD 8.0%),修复后降至26.0%(SD 4.4%)(P < 0.001)。
- 功能性重度模型平均反流分数为59.4%(SD 6.0%)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Transcatheter mitral valve repair is an emerging alternative to the surgical repair. This technology requires preclinical studies to assess efficacy in mitigating mitral regurgitation (MR). However, ex vivo MR models are not established. We developed 2 novel repairable models, functional and degenerative, which can quantitatively assess regurgitation and effect of intervention. METHODS: We used porcine mitral valves and a pulsatile flow circulation system. In the functional MR model, the annulus was immersed in 0.1% collagenase solution and dilated using 3D-printed dilators. To control the regurgitation grade, the sizes of the dilator and silicone sheet in which the valve was sutured to were adjusted. Chordae of P2 were severed in the degenerative model, and the number of severed chordae was adjusted to control the regurgitation grade. Models were repaired using the edge-to-edge or artificial chordae technique. RESULTS: The mean regurgitant fraction of the moderate-severe functional and degenerative models were 47.9% [standard deviation (SD): 2.2%] and 58.5% (SD: 8.0%), which were significantly reduced to 28.7% (SD: 4.4%) (P < 0.001) and 26.0% (SD: 4.4%) (P < 0.001) after the valve repair procedures. Severe functional model had a mean regurgitant fraction of 59.4% (SD: 6.0%). CONCLUSIONS: Both functional and degenerative models could produce sufficient MR levels that meet the interventional indication criteria. The repairable models are valuable in evaluating the efficacy of valve repair procedures and devices. The ability to control the amount of regurgitation enhances the versatility and reliability of these models. These reproducible models could expedite the development of novel devices.
引用
- Morimura H, Okamoto Y, Takada J, et al. Repairable ex vivo model of functional and degenerative mitral regurgitation. Eur J Cardiothorac Surg. 2023. PMID: 37951590. doi:10.1093/ejcts/ezad371