HeartMate 3 Snoopy:全磁悬浮血泵患者超声心动图转速阶梯测试与Valsalva动作中的无创心血管诊断

HeartMate 3 Snoopy: Noninvasive cardiovascular diagnosis of patients with fully magnetically levitated blood pumps during echocardiographic speed ramp tests and Valsalva maneuvers.
PubMed 文献 · J Heart Lung Transplant · 2023-09-26

这项前瞻性单中心研究用新型数据采集系统以1 Hz记录24例HeartMate 3(HM3)左心室辅助装置患者共35次转速阶梯测试(基线转速5443±244 rpm,变化范围±1000 rpm)的泵参数,并同步行超声心动图、心电图和血压测量。结果显示估计流量、电机电流、LVAD转速与血压呈正向共线性(r=0.9[0.1]),泵流量搏动性、搏动指数、转子噪声与左心室直径呈负向共线性(r=-0.8[0.1]),转子位移与心率无共线性(r=-0.1[0.08]);HM3 Snoopy可提供心-泵相互作用及搏动指数事件原因的诊断信息。

要点

  • HM3 Snoopy以1 Hz采集12项泵参数,同步超声、心电图和血压。
  • 估计流量、电机电流、LVAD转速与血压呈正向共线性(r=0.9[0.1])。
  • 泵流量搏动性、搏动指数、转子噪声与左心室直径呈负向共线性(r=-0.8[0.1])。

核心结论

  • HM3 Snoopy成功以更高采样率获取HM3更多参数,逐秒数据分析为心-泵相互作用及搏动指数事件原因提供额外临床诊断信息。

研究设计与样本

  • 前瞻性单中心研究,纳入24例HM3左心室辅助装置患者,共完成35次转速阶梯测试。
  • 用新型数据采集系统记录控制器与泵之间的HM3数据通信,每秒(1 Hz)采集12项泵参数。
  • 同步进行超声心动图、心电图和血压测量等临床评估,采用Pearson相关分析(r,中位数[IQR])。

主要发现

  • 患者年龄58.9±8.8岁,体重指数28.1±5.1 kg/m2,女性占20.8%。
  • 转速阶梯测试从基线转速5443±244 rpm起,变化范围为±1000 rpm。
  • 估计流量、电机电流、LVAD转速与血压共8项HM3泵参数呈正向共线性(r=0.9[0.1])。
  • 泵流量搏动性、搏动指数、转子噪声与左心室直径呈负向共线性(r=-0.8[0.1])。
  • 转子位移与心率无共线性(r=-0.1[0.08])。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The HeartMate 3 (HM3) left ventricular assist device (LVAD) has demonstrated excellent clinical outcomes; however, pump speed optimization is challenging with the available HM3 monitoring. Therefore, this study reports on clinical HM3 parameters collected with a noninvasive HM3 monitoring system (HM3 Snoopy) during echocardiographic speed ramp tests and Valsalva maneuvers. METHODS: In this prospective, single-center study, the HM3 data communication between the controller and pump was recorded with a novel data acquisition system. Twelve pump parameters sampled every second (1 Hz) and clinical assessments (echocardiography, electrocardiogram (ECG), and blood pressure measurement) during speed ramp tests were analyzed using Pearson's correlation (r, median [IQR]). The cause for the occurrence of pulsatility index (PI)-events during ramp speed tests and valsalva maneuvers was investigated. RESULTS: In 24 patients (age: 58.9 ± 8.8 years, body mass index: 28.1 ± 5.1 kg/m2, female: 20.8%), 35 speed ramp tests were performed with speed changes in the range of ±1000 rpm from a baseline speed of 5443 ± 244 rpm. Eight HM3 pump parameters from estimated flow, motor current, and LVAD speed together with blood pressure showed positive collinearities (r = 0.9 [0.1]). Negative collinearities were observed for pump flow pulsatility, pulsatility index, rotor noise, and left ventricular diameters (r = -0.8 [0.1]), whereas rotor displacement and heartrate showed absence of collinearities (r = -0.1 [0.08]). CONCLUSIONS: In this study, the HM3 Snoopy was successfully used to acquire more parameters from the HM3 at a higher sampling rate. Analysis of HM3 per-second data provide additional clinical diagnostic information on heart-pump interactions and cause of PI-events.

来源:J Heart Lung Transplant,查看原文。版权归原作者所有。

引用

  1. Schlöglhofer T, Gross C, Abart T, et al. HeartMate 3 Snoopy: Noninvasive cardiovascular diagnosis of patients with fully magnetically levitated blood pumps during echocardiographic speed ramp tests and Valsalva maneuvers. J Heart Lung Transplant. 2023. PMID: 37769887. doi:10.1016/j.healun.2023.09.011