体外驱动线振动检测左心室辅助装置血栓形成——猪模型研究

Extracorporeal driveline vibrations to detect left ventricular assist device thrombosis - A porcine model study.
PubMed 文献 · J Heart Lung Transplant · 2023-09-04

在6只植入HeartWare心室辅助装置(HVAD)的猪模型中,将加速度计分别贴于泵体和驱动线,共注射59次血栓以诱导泵血栓形成(PT),比较两处信号的频谱图与第三谐波显著度(S3H)。5次实验驱动线信号良好,驱动线第三谐波可见20次而参考为30次,S3H强相关且曲线下面积(AUC)为0.85。作者认为体外驱动线加速度计是HVAD泵血栓检测的可行方法。

要点

  • 猪HVAD模型中驱动线与泵体加速度计信号在频谱图上关系清晰。
  • 驱动线第三谐波可见20次,参考加速度计为30次。
  • S3H相关性强,检测泵血栓的AUC为0.85。

核心结论

  • 体外驱动线加速度计可在加速度计与左心室辅助装置整合可行之前,作为准确检测HVAD泵血栓的现成方法。

研究设计与样本

  • 猪HVAD模型(n=6),将加速度计分别贴于泵体作参考和驱动线作比较,共向心脏注射59次血栓诱导泵血栓形成,随后进行间歇血栓冲洗操作。

主要发现

  • 5次实验驱动线信号强度良好,两加速度计频谱图关系清晰。
  • 驱动线第三谐波可见20次,参考加速度计为30次。
  • S3H比较显示强相关,AUC为0.85。
  • S3H在噪声和假性泵血栓检测方面表现稳健。

关键数值与推荐

  • 驱动线第三谐波可见20次 vs 参考30次。
  • S3H相关分析AUC为0.85。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Pump thrombosis (PT) and related adverse complications contributed to the HeartWare Ventricular Assist Device (HVAD) market withdrawal. Many patients still receive lifelong support, with deficient PT surveillance based on pump power trends. Analysis of pump vibrations is better for detecting PT. Here, we investigated the feasibility of an extracorporeal accelerometer to detect PT from pump vibrations propagated out on the driveline. METHODS: In a porcine HVAD model (n = 6), an accelerometer was attached to the pump as a reference and another to the driveline for comparisons of signals. In total, 59 thrombi were injected into the heart to induce PT, followed by intermittent thrombus washout maneuvers. Signals were compared visually in spectrograms and quantitatively in third harmonic saliences (S3H) by correlation analysis. Receiver operating characteristic curves expressed the method's outcome in sensitivity vs specificity, with the overall diagnostic performance in the area under the curve (AUC) score. RESULTS: Five experiments had good driveline signal strength, with clear spectrographic relationships between the 2 accelerometers. Third harmonic driveline vibrations were visible 20 vs 30 times in the reference. The comparison in S3H showed a strong correlation and yielded an AUC of 0.85. Notably, S3H proved robust regarding noise and false PT detections. CONCLUSIONS: An extracorporeal accelerometer on the driveline can be a readily available method for accurate HVAD PT detection before an accelerometer integration with left ventricular assist device is feasible.