文献速递先天性心脏病外科

用于儿童心脏外科手术补片制备的激光投影平台验证

Validation of a laser projection platform for the preparation of surgical patches used in paediatric cardiac surgery.
PubMed 文献 · Interdiscip Cardiovasc Thorac Surg · 2023-01-01

本研究验证MicroLASERGUIDE激光投影平台用于制备儿童心脏手术补片的可行性。以数字版OZAKI AVNeo模板尺寸为对照,比较激光投影与物理模板法制备51个13号牛心包瓣叶的尺寸差异,并测试其他补片材料和尺寸的可行性。激光组尺寸与对照接近,最大偏差约5%(约1毫米),展示出作为手术补片制备替代方法的潜力。

要点

  • 激光投影组制备的瓣叶尺寸与数字对照接近,高度差0.0毫米,长度差0.2毫米。
  • 物理模板组高度和长度均小于对照,差值分别为-0.3毫米和-0.4毫米。
  • 其他补片材料和尺寸测试结果相似,激光组尺寸与对照最大偏差约5%(约1毫米)。

核心结论

  • 激光投影平台作为心脏外科手术补片制备的替代方法展示出前景。
  • 该技术有潜力革新需要患者特异性补片修补的多种先天性异常的术前手术规划。

研究设计与样本

  • 采用非劣效性比较,在未固定牛心包上制备51个13号瓣叶,激光投影组与OZAKI AVNeo模板组各51个,以数字版OZAKI AVNeo模板尺寸为对照。
  • 可行性测试使用其他常见补片材料(固定牛心包、PTFE和猪主肺动脉作为肺动脉同种移植物的替代)和尺寸(13、23),每组n=3。

主要发现

  • 与对照(高度21.5毫米,长度21.0毫米)相比,模板组高度和长度较小,高度差-0.3(-0.5至0.0)毫米,长度差-0.4(-0.8至-0.1)毫米,P均<0.01。
  • 激光组高度和长度与对照相对相似,高度差0.0(-0.2至0.2)毫米,P=0.804;长度差0.2(-0.1至0.4)毫米,P=0.029。
  • 模板组高度和长度百分比误差分别为-1.5(-2.3至0.0)%和-1.9(-3.7至-0.6)%,激光组分别为0.2(-1.0至1.1)%和1.0(-0.5至1.8)%。
  • 其他材料和尺寸测试结果相似;总体而言,激光组样本尺寸与对照最大偏差为5%(约1毫米)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Reconstruction of cardiovascular anatomy with patch material is integral to the repair of congenital heart disease. We present validation of a laser projection platform for the preparation of surgical patches as a proof-of-concept for intraoperative use in patient-specific planning of paediatric cardiac surgery reconstructions. METHODS: The MicroLASERGUIDE, a compact laser projection system that displays computer-aided designs onto 2D/3D surfaces, serves as an alternative to physical templates. A non-inferiority comparison of dimensional measurements was conducted between laser projection ('laser') and OZAKI AVNeo Template ('template') methods in creation of 51 (each group) size 13 valve leaflets from unfixed bovine pericardium. A digital version of the OZAKI AVNeo Template dimensions served as control. Feasibility testing was performed with other common patch materials (fixed bovine pericardium, PTFE and porcine main pulmonary artery as a substitute for pulmonary homograft) and sizes (13, 23) (n = 3 each group). RESULTS: Compared to control (height 21.5, length 21.0 mm), template height and length were smaller (height and length differences of -0.3 [-0.5 to 0.0] and -0.4 [-0.8 to -0.1] mm, P < 0.01 each); whereas, both laser height and length were relatively similar (height and length differences of height 0.0 [-0.2 to 0.2], P = 0.804, and 0.2 [-0.1 to 0.4] mm, P = 0.029). Template percent error for height and length was -1.5 (-2.3 to 0.0)% and -1.9 (-3.7 to -0.6)% vs 0.2 (-1.0 to 1.1)% and 1.0 (-0.5 to 1.8)% for the laser. Similar results were found with other materials and sizes. Overall, laser sample dimensions differed by a maximum of 5% (∼1 mm) from the control. CONCLUSIONS: The laser projection platform has demonstrated promise as an alternative methodology for the preparation of surgical patches for use in cardiac surgery. This technology has potential to revolutionize preoperative surgical planning for numerous congenital anomalies that require patient-specific patch-augmented repair.

来源:Interdiscip Cardiovasc Thorac Surg,查看原文。版权归原作者所有。

引用

  1. Saunders T, Recco D, Kneier N, et al. Validation of a laser projection platform for the preparation of surgical patches used in paediatric cardiac surgery. Interdiscip Cardiovasc Thorac Surg. 2023. PMID: 37555820. doi:10.1093/icvts/ivad129

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。