主动脉瓣可视化和加压装置:一种用于术中评估主动脉瓣修复手术的新型装置
本研究开发了一种新型术中主动脉瓣可视化和加压(AVP)装置,可在心脏停搏下对修复后的主动脉瓣进行生理条件评估并测量瓣膜反流量。在24例保留瓣膜的主动脉根部置换患者中使用该装置,术后22例超声心动图主动脉瓣反流≤1级,中位漏量为90 ml/min(IQR 60-120 ml/min),3例根据直视检查行额外调整,2例复杂解剖患者行瓣膜置换。作者认为该装置可使术中评估更可预测、手术更高效。
要点
- AVP装置可在心脏停搏下对修复后主动脉瓣加压至60-80 mmHg并内镜直视评估。
- 24例保留瓣膜根部置换患者中,22例术后超声主动脉瓣反流≤1级。
- 中位漏量为90 ml/min(IQR 60-120 ml/min),3例据直视检查行额外调整。
核心结论
- 新型AVP装置可在心脏停搏下于生理条件下术中评估主动脉瓣,并允许针对性调整。
- 作者认为该装置可作为主动脉瓣修复和保留瓣膜手术术中评估的重要辅助,使手术结果更可预测、手术更高效。
研究设计与样本
- 单臂器械开发与术中应用研究,在24例接受保留瓣膜的主动脉根部置换患者中使用AVP装置。
- 装置连接于(新)主动脉,心脏停搏下以盐水加压根部至60-80 mmHg并置入内镜,直视检查瓣膜并测量瓣膜漏量。
- 术后以经食管超声心动图测量主动脉瓣反流作为金标准,与装置测量的反流量进行比较。
主要发现
- 24例患者中,22例术后超声心动图主动脉瓣反流≤1级。
- 中位漏量为90 ml/min(IQR 60-120 ml/min)。
- 3例患者在直视检查后行额外调整。
- 2例复杂解剖患者行瓣膜置换;其中1例装置评估后直视结果不理想且残余反流330 ml/min,另1例测得残余反流260 ml/min且直视检查见瓣膜受限。
临床怎么用
- 作者认为AVP装置可作为主动脉瓣修复和保留瓣膜手术术中评估的重要辅助。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Aortic valve repair procedures are technically challenging, and current intraoperative evaluation methods often fail to predict the final echocardiographic result. We have developed a novel intraoperative aortic valve visualization and pressurization (AVP) device, enabling valve inspection under physiological conditions, and measuring aortic valve insufficiency (AI) during cardioplegic arrest. METHODS: The AVP device is attached to the (neo)aorta, after any type of aortic valve repair, while the heart is arrested. The root is pressurized (60-80 mmHg) using a saline solution and an endoscope is introduced. The valve is inspected, and the amount of valvular leakage is measured. Postoperative 'gold standard' transesophageal echocardiogram measurements of AI are performed and compared against regurgitation volume measured. RESULTS: In 24 patients undergoing valve-sparing root replacement, the AVP device was used. In 22 patients, postoperative echocardiographic AI was ≤ grade 1. The median leakage was 90 ml/min, IQR 60-120 ml/min. In 3 patients, additional adjustments after visual inspection was performed. In 2 patients, with complex anatomy, the valve was replaced. In one, after evaluation with the device, there was undesirable result visually and residual AI of 330 ml/min, and in another, 260 ml/min residual AI was measured and valve restriction on visual inspection. CONCLUSIONS: The novel AVP device enables intraoperative evaluation of the valve under physiological conditions, while still on arrested heart, and allows for targeted adjustments. The AVP device can be an important aid for intraoperative evaluation of the aortic valve, during valve repair and valve-sparing procedures, thereby making the operative result more predictable and the operation more efficient.
引用
- Arabkhani B, Sandker SC, Braun J, et al. Aortic valve visualization and pressurization device: a novel device for intraoperative evaluation of aortic valve repair procedures. Eur J Cardiothorac Surg. 2023. PMID: 37610333. doi:10.1093/ejcts/ezad291