连续流左心室辅助装置患者大循环与微循环动脉的搏动性与血流模式

Pulsatility and flow patterns across macro- and microcirculatory arteries of continuous-flow left ventricular assist device patients.
PubMed 文献 · J Heart Lung Transplant · 2023-04-23

该研究采用二维角度校正多普勒超声,对148名参与者(健康对照32例、心力衰竭43例、HeartMate II 32例、HeartMate3 41例)的颈总动脉、大脑中动脉和视网膜中央动脉搏动指数进行量化。结果显示HeartMate3的「人工脉冲」在大循环和微循环中可检测到,但与HeartMate II相比未产生搏动指数的显著改变;泵速与微循环搏动指数呈负相关。

要点

  • HeartMate3「人工脉冲」搏动与连续流搏动的大循环和微循环搏动指数与HeartMate II相似。
  • HeartMate3与HeartMate II的收缩期峰值流速无差异。
  • HeartMate3和HeartMate II的搏动指数向微循环传递均高于心力衰竭患者。

核心结论

  • HeartMate3的「人工脉冲」在大循环和微循环中可检测到,但与HeartMate II患者相比未产生搏动指数的显著改变。
  • 搏动性向微循环传递增加及泵速与微循环搏动指数的关联提示,未来HeartMate3患者的临床照护可能涉及根据特定终末器官的微循环搏动指数个体化调整泵设置。

研究设计与样本

  • 采用二维对齐、角度校正的多普勒超声量化颈总动脉、大脑中动脉和视网膜中央动脉(代表微循环)的局部血流振荡(搏动指数)。
  • 共纳入148名参与者:健康对照32例、心力衰竭43例、HeartMate II 32例、HeartMate3 41例。

主要发现

  • HeartMate3患者「人工脉冲」搏动与「连续流」搏动的二维多普勒搏动指数在大循环和微循环中与HeartMate II患者相似。
  • HeartMate3与HeartMate II患者的收缩期峰值流速无差异。
  • HeartMate3(「人工脉冲」搏动期间)和HeartMate II患者的搏动指数向微循环传递均高于心力衰竭患者。
  • 左心室辅助装置泵速与微血管搏动指数呈负相关:HeartMate II r²=0.51,p<0.0001;HeartMate3「连续流」r²=0.32,p=0.0009;HeartMate3「人工脉冲」r²=0.23,p=0.007。
  • 左心室辅助装置泵搏动指数仅与HeartMate II患者的微循环搏动指数相关。

临床怎么用

  • 原文结论指出,未来HeartMate3患者的临床照护可能涉及根据特定终末器官的微循环搏动指数个体化调整泵设置。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Reduced arterial pulsatility in continuous-flow left ventricular assist devices (CF-LVAD) patients has been implicated in clinical complications. Consequently, recent improvements in clinical outcomes have been attributed to the "artificial pulse" technology inherent to the HeartMate3 (HM3) LVAD. However, the effect of the "artificial pulse" on arterial flow, transmission of pulsatility into the microcirculation and its association with LVAD pump parameters is not known. METHODS: The local flow oscillation (pulsatility index, PI) of common carotid arteries (CCAs), middle cerebral arteries (MCAs) and central retinal arteries (CRAs-representing the microcirculation) were quantified by 2D-aligned, angle-corrected Doppler ultrasound in 148 participants: healthy controls, n = 32; heart failure (HF), n = 43; HeartMate II (HMII), n = 32; HM3, n = 41. RESULTS: In HM3 patients, 2D-Doppler PI in beats with "artificial pulse" and beats with "continuous-flow" was similar to that of HMII patients across the macro- and microcirculation. Additionally, peak systolic velocity did not differ between HM3 and HMII patients. Transmission of PI into the microcirculation was higher in both HM3 (during the beats with "artificial pulse") and in HMII patients compared with HF patients. LVAD pump speed was inversely associated with microvascular PI in HMII and HM3 (HMII, r2 = 0.51, p < 0.0001; HM3 "continuous-flow," r2 = 0.32, p = 0.0009; HM3 "artificial pulse," r2 = 0.23, p = 0.007), while LVAD pump PI was only associated with microcirculatory PI in HMII patients. CONCLUSIONS: The "artificial pulse" of the HM3 is detectable in the macro- and microcirculation but without creating a significant alteration in PI compared with HMII patients. Increased transmission of pulsatility and the association between pump speed and PI in the microcirculation indicate that the future clinical care of HM3 patients may involve individualized pump settings according to the microcirculatory PI in specific end-organs.