加速度计检测左心室辅助装置流入道梗阻:猪模型研究

Detection of inflow obstruction in left ventricular assist devices by accelerometer: A porcine model study.
PubMed 文献 · J Heart Lung Transplant · 2023-04-05

在8只猪模型中,用球囊导管将HVAD流入道阻塞34%至94%共5个级别,发现非谐波振幅(NHA)在阻塞52%-83%时升高,而泵功率(PLVAD)变化较小;NHA的AUC为0.85-1.00,PLVAD为0.35-0.73。结论认为NHA升高可可靠提示亚临床渐进性流入道梗阻。

要点

  • 猪模型(n=8)中,流入道阻塞52%-83%时NHA升高,PLVAD变化较小。
  • NHA的AUC为0.85-1.00,PLVAD的AUC为0.35-0.73。
  • 加速度计可能补充PLVAD用于更早预警和泵定位。

核心结论

  • NHA升高可可靠提示亚临床渐进性流入道梗阻,加速度计可能补充PLVAD用于更早预警和泵定位。

研究设计与样本

  • 猪模型研究(n=8),用球囊导管将HVAD流入道阻塞34%至94%共5个级别,并以增加后负荷和改变转速作为对照。

主要发现

  • 对照干预期间NHA仅受轻微影响,而PLVAD不同。
  • 阻塞52%-83%期间NHA升高,其中质量摆动最明显,同时PLVAD变化远较小。
  • 增加泵转速倾向于放大NHA升高。
  • NHA的AUC为0.85-1.00,PLVAD的AUC为0.35-0.73。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Left ventricular assist devices (LVAD) provide circulatory blood pump support for severe heart failure patients. Pump inflow obstructions may lead to stroke and pump malfunction. We aimed to verify in vivo that gradual inflow obstructions, representing prepump thrombosis, are detectable by a pump-attached accelerometer, where the routine use of pump power (PLVAD) is deficient. METHOD: In a porcine model (n = 8), balloon-tipped catheters obstructed HVAD inflow conduits by 34% to 94% in 5 levels. Afterload increases and speed alterations were conducted as controls. We computed nonharmonic amplitudes (NHA) of pump vibrations captured by the accelerometer for the analysis. Changes in NHA and PLVAD were tested by a pairwise nonparametric statistical test. Detection sensitivities and specificities were investigated by receiver operating characteristics with areas under the curves (AUC). RESULTS: NHA remained marginally affected during control interventions, unlike PLVAD. NHA elevated during obstructions within 52-83%, while mass pendulation was most pronounced. Meanwhile, PLVAD changed far less. Increased pump speeds tended to amplify the NHA elevations. The corresponding AUC was 0.85-1.00 for NHA and 0.35-0.73 for PLVAD. CONCLUSION: Elevated NHA provides a reliable indication of subclinical gradual inflow obstructions. The accelerometer can potentially supplement PLVAD for earlier warnings and localization of pump.

来源:J Heart Lung Transplant,查看原文。版权归原作者所有。