文献速递心外重症与机械循环支持

左心室辅助装置模式:在离体心脏灌注工作模式中协同脉冲左心室卸载

Left ventricular assist device mode: Co-pulse left ventricular unloading in a working mode of ex vivo heart perfusion.
PubMed 文献 · J Heart Lung Transplant · 2023-02-02

研究在常温离体心脏灌注中比较静息模式、工作模式与新型左心室辅助装置(LVAD)协同脉冲模式对猪心脏(n=18)心肌功能的保护效果。LVAD组在心输出量、每搏功、dP/dt最大值和最小值等保存率上显著高于其他两组,且心肌氧耗和ATP水平更高。数据提示LVAD模式的心肌机械功能保存优于静息和工作模式。

要点

  • LVAD协同脉冲模式在离体心脏灌注中显著提高心肌机械功能保存率。
  • LVAD组心输出量保存率76%±5%,高于静息组33±3和工作组35±5(p<0.001)。
  • LVAD组ATP水平1.6±0.2 μmol/g,高于静息组1.1±0.1和工作组0.7±0.1(p=0.001)。

核心结论

  • 数据提示LVAD模式的心肌机械功能保存优于静息模式和工作模式。
  • LVAD组和工作组糖酵解活性相似,LVAD组最终ATP更高可能归因于LVAD减少外部做功。

研究设计与样本

  • 猪心脏(n=18)在0至1小时采用静息模式灌注,1至5小时分别采用静息、工作或LVAD模式灌注,6小时在工作模式下评估心肌功能。
  • 保存率定义为330分钟时心肌机械功能除以75分钟时功能。

主要发现

  • LVAD组心输出量保存率76%±5%,静息组33±3,工作组35±5,LVAD组显著更高(p<0.001)。
  • LVAD组每搏功、dP/dt最大值和dP/dt最小值的保存率显著高于其他两组。
  • 静息组葡萄糖消耗显著降低。
  • LVAD组心肌氧耗6.1±0.5 mL O2/min/100 g,静息组2.2±0.3,工作组4.6±0.5,LVAD组显著更高(p<0.001)。
  • LVAD组ATP水平1.6±0.2 μmol/g,静息组1.1±0.1,工作组0.7±0.1,LVAD组显著更高(p=0.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: For normothermic ex vivo heart perfusion (EVHP), a resting mode and working mode have been proposed. We newly developed a left ventricular assist device (LVAD) mode that supports heart contraction by co-pulse synchronized LVAD. METHODS: Following resting mode during time 0 to 1 hour, pig hearts (n = 18) were perfused in either resting, working, or LVAD mode during time 1 to 5 hour, and then myocardial function was evaluated in working mode at 6 hour. The preservation ratio was defined as the myocardial mechanical function at 330 minute divided by the function at 75 minute. In LVAD mode, LVAD unloaded the pressure and the volume in the left ventricle in the systolic phase. RESULTS: The LVAD group was significantly associated with higher preservation ratios in cardiac output (resting, 33 ± 3; working, 35 ± 5; LVAD, 76% ± 5%; p < 0.001), stroke work, dP/dt maximum, and dP/dt minimum compared with the other groups. Glucose consumption was significantly reduced in the resting group. The LVAD group was significantly associated with higher myocardial oxygen consumption (resting, 2.2 ± 0.3; working; 4.6 ± 0.5; LVAD, 6.1 ± 0.5 mL O2/min/100 g, p < 0.001) and higher adenosine triphosphate (ATP) levels (resting, 1.1 ± 0.1; working, 0.7 ± 0.1; LVAD, 1.6 ± 0.2 μmol/g, p = 0.001) compared with the others. CONCLUSION: These data suggest that myocardial mechanical function was better preserved in LVAD mode than in resting and working modes. Although our data suggested similar glycolysis activity in the LVAD and working groups, the higher final ATP in the LVAD group might be explained by reduced external work in LVAD.