左心室辅助装置流入道插管对位不佳导致反复低流量报警
这项回顾性研究分析了48例HeartMate 3左心室辅助装置支持患者,采用对比增强CT标准化方案测量流入道插管角度,使用gamma脆弱模型分析角度与反复低流量报警的相关性。结果显示朝向室间隔-侧壁平面成角与低流量报警随时间增加显著相关(P=0.048),室间隔-侧壁成角≥28°时报警显著增加(P=0.001)。
要点
- 朝向室间隔-侧壁平面每增加1度,随访每月低流量报警以恒定速率函数增加0.031(P=0.048)。
- 室间隔-侧壁成角≥28°时低流量报警显著增加(P=0.001)。
- 前后向与最大流入道插管成角未见显著差异。
核心结论
- 左心室辅助装置流入道插管朝向室间隔-侧壁平面成角增大时低流量报警次数增加,强调流入道插管角度位置对预防反复低流量报警的重要性。
研究设计与样本
- 回顾性研究,纳入48例HeartMate 3左心室辅助装置支持患者,建立标准化方案用对比增强计算机断层扫描测量流入道插管角度,采用恒定速率函数的gamma脆弱模型进行统计。
主要发现
- 48例患者中男性占79%,中位年龄57岁,中位随访30个月(四分位距19-41)。
- 30例(63%)患者出现低流量报警。
- 朝向室间隔-侧壁平面成角与低流量报警随时间增加显著相关,恒定速率函数为随访每月每增加1度成角低流量报警增加0.031(P=0.048)。
- 以最佳截断点分组,室间隔-侧壁成角≥28°时低流量报警显著增加(P=0.001)。
- 前后向与最大流入道插管成角未见显著差异。
临床怎么用
- 原文结论强调左心室辅助装置流入道插管角度位置对预防反复低流量报警的重要性,原文未据此给出具体推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms. METHODS: Medical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized protocol was created to measure the angular position with a contrast-enhanced computed tomography scan. Statistics were done using a gamma frailty model with a constant rate function. RESULTS: For this analysis, 48 LVAD-supported patients were included. The majority of the patients were male (79%) with a median age of 57 years and a median follow-up of 30 months (interquartile range: 19-41). Low-flow alarm(s) were experienced in 30 (63%) patients. Angulation towards the septal-lateral plane showed a significant increase in low-flow alarms over time with a constant rate function of 0.031 increase in low-flow alarms per month of follow-up per increasing degree of angulation (P = 0.048). When dividing this group using an optimal cut-off point, a significant increase in low-flow alarms was observed when the septal-lateral angulation was 28° or more (P = 0.001). Anterior-posterior and maximal inflow cannula angulation did not show a significant difference. CONCLUSIONS: This study showed an increasing number of low-flow alarms when the degrees of LVAD inflow cannula expand towards the septal-lateral plane. This emphasizes the importance of the LVAD inflow cannula angular position to prevent relapsing low-flow alarms with the risk of diminished quality of life and morbidity.
引用
- Zijderhand CF, Knol WG, Budde RPJ, et al. Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula. Eur J Cardiothorac Surg. 2022. PMID: 35993906. doi:10.1093/ejcts/ezac415
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。