文献速递心外重症与机械循环支持

心室辅助装置导线皮肤整合的体内评估

In vivo evaluation of skin integration with ventricular assist device drivelines.
PubMed 文献 · J Heart Lung Transplant · 2022-03-29

本研究在42只大鼠背部经皮植入Heartmate 3导线片段,比较光滑聚氨酯、丝绒和熔融静电纺丝(MEW)支架三种表面的皮肤整合。结果显示丝绒和MEW支架促进表皮整合,MEW组组织长入显著高于丝绒组(7天和14天均p<0.0001),而光滑聚氨酯无附着并出现袋状化。

要点

  • 丝绒和MEW支架表面在导线-皮肤界面锚定,光滑聚氨酯无附着。
  • MEW组组织长入在7天和14天均显著高于丝绒组(p<0.0001)。
  • 丝绒与MEW支架组的机械拔出力在7天和14天相似(p>0.05)。

核心结论

  • 丝绒和MEW支架促进表皮整合,而光滑聚氨酯导线未促进表皮整合。
  • 精细控制MEW支架结构生产实现完全细胞覆盖,可能减少导线感染。

研究设计与样本

  • 在42只大鼠背部经皮植入导线片段,每只大鼠植入2根2 cm长的导线。
  • 比较光滑聚氨酯、丝绒 bonded to 光滑聚氨酯、以及带MEW支架套筒的光滑聚氨酯三种表面。
  • 在7天和14天后评估皮肤整合。

主要发现

  • 宏观上,丝绒和MEW支架表面在导线-皮肤界面锚定,而光滑聚氨酯样品无附着。
  • 组织学分析显示表皮迁移贯穿丝绒和MEW支架组的厚度。
  • MEW组组织长入在7天(p<0.0001)和14天(p<0.0001)显著高于丝绒组。
  • 光滑聚氨酯样品中观察到袋状化。
  • 丝绒与MEW支架组的机械拔出力在7天和14天相似(p>0.05)。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Ventricular assist device (VAD) driveline exit site infection is a common complication. 3D scaffolds manufactured with highly homogeneous pores via melt electro-writing (MEW) may generate an improved skin-driveline interface which permits cellular in-growth and creates a barrier to prevent bacterial migration along the driveline tissue tunnel. This study investigated skin integration on segments of Heartmate 3 driveline: smooth polyurethane, velour, and on a custom MEW scaffold in a small animal model. METHODS: Drivelines with surfaces consisting of smooth polyurethane, velour bonded to smooth polyurethane, and smooth polyurethane with a MEW scaffold sleeve were implanted percutaneously in the dorsum of 42 rats. Each rat was implanted with 2 pieces of driveline of 2 cm in length. Skin integration was assessed after 7 and 14 days. RESULTS: Macroscopically, velour and MEW scaffold surfaces were anchored at the driveline-skin interface while smooth polyurethane samples were not attached. The histology analyses showed epidermal migration throughout the thickness of the velour and MEW scaffold groups. Evident tissue growth around single MEW scaffold fibers resulted in full coverage of the pores, while areas of compacted fibers were apparent in the velour group. Tissue ingrowth was significantly higher in the MEW group compared to the velour group after 7 (p < 0.0001) and 14 days (p < 0.0001). Marsupialization was observed in the smooth polyurethane samples. Mechanical pull-out forces were similar between velour and MEW scaffold groups at 7 and 14 days (p > 0.05). CONCLUSIONS: Velour and MEW scaffolds promoted epidermal integration while smooth polyurethane drivelines did not. Fine control of MEW scaffold structure production resulted in full cellular coverage and may reduce driveline infection.

来源:J Heart Lung Transplant,查看原文。版权归原作者所有。

引用

  1. Cavalcanti AS, Diaz RS, Bolle ECL, et al. In vivo evaluation of skin integration with ventricular assist device drivelines. J Heart Lung Transplant. 2022. PMID: 35641425. doi:10.1016/j.healun.2022.03.014

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。