自动钛紧固件对主动脉瓣生物瓣的蛀蚀样影响:一则警示
一例78岁患者在外院行第三次生物主动脉瓣置换(25 mm Inspiris Resilia)时使用COR-KNOT紧固件,2021年因重度主动脉瓣反流入院行第四次手术。术中见全部3个瓣叶共7处穿孔,明显由刚性金属紧固件所致。作者认为COR-KNOT所致缺损在术后4个月内出现,此类严重并发症的真实发生率可能被低估,建议在生物主动脉瓣置换中慎用COR-KNOT并密切术后随访。
要点
- COR-KNOT所致瓣叶穿孔并非术后立即出现,而是在术后4个月内出现。
- 因主动脉瓣反流导致的生物瓣失败常被归为「早期退化」,真实发生率可能被低估。
- 作者建议在生物主动脉瓣置换中慎用COR-KNOT,并密切术后随访。
核心结论
- COR-KNOT所致缺损并非术后立即出现,而是在术后4个月内出现;此类严重并发症的真实发生率可能被低估。
- 作者认为在生物主动脉瓣置换中应慎用COR-KNOT,患者应接受密切的术后随访。
研究设计与样本
- 病例报告:一例78岁患者在外院行第三次生物主动脉瓣置换(25 mm,Inspiris Resilia,Edwards Lifesciences LLC,Irvine,CA,USA),因瓣膜退化使用COR-KNOT(LSI Solutions,Victor,NY,USA)。
- 2021年患者被诊断为重度主动脉瓣反流,入住本院接受第四次手术。
主要发现
- 术中观察到全部3个瓣叶共7处穿孔,明显由刚性金属紧固件所致。
- 行再次孤立性主动脉瓣置换(29 mm,Perimount Magna Ease,Edwards Lifesciences LLC),采用常规打结、带垫片编织线。
- 术后病程平稳。
临床怎么用
- 作者认为,在生物主动脉瓣置换中应慎用COR-KNOT,患者应接受密切的术后随访。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
A 78-year-old patient underwent his third biological aortic valve replacement (25 mm, Inspiris Resilia™, Edwards Lifesciences LLC, Irvine, CA, USA) with COR-KNOT® (LSI Solutions, Victor, NY, USA) due to valve degeneration in 2018 at a foreign hospital. In 2021, the patient was diagnosed with severe aortic regurgitation and admitted to our hospital for his fourth surgery. Intraoperatively, a total of 7 perforations in all 3 valve leaflets have been observed, which were obviously induced by the rigid metallic fasteners. Redo isolated aortic valve replacement (29 mm, Perimount Magna Ease, Edwards Lifesciences LLC) was performed with conventionally knotted, pledget enforced braided threads. Postoperative course was uneventful. This report shows that COR-KNOT-induced defects do not appear immediately after surgery but within the first 4 postoperative months. Since failure of aortic bioprostheses due to aortic insufficiency is often rated as 'early degeneration' and degenerated aortic bioprostheses are mostly replaced by valve-in-valve strategy, the true incidence of this serious complication might be underestimated. COR-KNOT should be used with caution in biological aortic valve replacement and patients should undergo close postoperative follow-ups.
引用
- Santer D, Miazza J, Eckstein F, et al. Moth-eaten like impact of automated titanium fasteners on aortic valve bioprosthesis: a word of caution. Eur J Cardiothorac Surg. 2022. PMID: 35758627. doi:10.1093/ejcts/ezac357