文献速递心外重症与机械循环支持

一种新型心包内搏动性装置用于个体化双心室循环支持且不直接接触血液

A novel intrapericardial pulsatile device for individualized, biventricular circulatory support without direct blood contact.
PubMed 文献 · J Thorac Cardiovasc Surg · 2022-03-01

研究评估新型微创机械循环支持装置reBEAT的功能与安全性,该装置完全避免血液接触并提供搏动性双心室支持。在7头猪中(50-80 kg),急性期(β受体阻滞下,n=5)和30天长时生存模型(n=2)分析显示装置成功放置,心电图信号自动标注实现实时机械支持同步,急性动物血流动力学参数逐步改善,长时生存显示装置整合安全且心电图信号检测清晰稳定。

要点

  • reBEAT装置完全避免血液接触,提供搏动性双心室支持。
  • 在7头猪中,装置均成功放置于心室周围。
  • 心电图信号自动标注实现实时机械支持同步,急性动物血流动力学逐步改善。

核心结论

  • 本研究证明reBEAT可提供双心室心脏支持,其安全植入、解剖适配、装置-组织整合安全及实时心电图同步机械支持等特性对临床转化至关重要,并实现有效装置功能和长期安全性。

研究设计与样本

  • 在7头猪(体重50-80 kg)中分析急性期数据(β受体阻滞下,n=5)和30天长时生存模型(n=2)。
  • 每只动物基于计算机断层扫描分析制造尺寸精确的心脏植入物。

主要发现

  • 每个植入物均成功放置于心室周围。
  • 自动算法心电图信号标注实现与每个心动周期精确的实时机械支持同步。
  • 急性动物心脏血流动力学参数逐步改善。
  • 长时生存显示装置整合安全,且心电图信号检测随时间清晰稳定。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Due to severely limited donor heart availability, durable mechanical circulatory support remains the only treatment option for many patients with end-stage heart failure. However, treatment complexity persists due to its univentricular support modality and continuous contact with blood. We investigated the function and safety of reBEAT (AdjuCor GmbH), a novel, minimal invasive mechanical circulatory support device that completely avoids blood contact and provides pulsatile, biventricular support. METHODS: For each animal tested, an accurately sized cardiac implant was manufactured from computed tomography scan analyses. The implant consists of a cardiac sleeve with three inflatable cushions, 6 epicardial electrodes and driveline connecting to an electro-pneumatic, extracorporeal portable driver. Continuous epicardial electrocardiogram signal analysis allows for systolic and diastolic synchronization of biventricular mechanical support. In 7 pigs (weight, 50-80 kg), data were analyzed acutely (under beta-blockade, n = 5) and in a 30-day long-term survival model (n = 2). Acquisition of intracardiac pressures and aortic and pulmonary flow data were used to determine left ventricle and right ventricle stroke work and stroke volume, respectively. RESULTS: Each implant was successfully positioned around the ventricles. Automatic algorithm electrocardiogram signal annotations resulted in precise, real-time mechanical support synchronization with each cardiac cycle. Consequently, progressive improvements in cardiac hemodynamic parameters in acute animals were achieved. Long-term survival demonstrated safe device integration, and clear and stable electrocardiogram signal detection over time. CONCLUSIONS: The present study demonstrates biventricular cardiac support with reBEAT. Various demonstrated features are essential for realistic translation into the clinical setting, including safe implantation, anatomical fit, safe device-tissue integration, and real-time electrocardiogram synchronized mechanical support, result in effective device function and long-term safety.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Schueler S, Bowles CT, Hinkel R, et al. A novel intrapericardial pulsatile device for individualized, biventricular circulatory support without direct blood contact. J Thorac Cardiovasc Surg. 2022. PMID: 35379474. doi:10.1016/j.jtcvs.2021.11.093

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。