本研究在2018年11月至2020年12月间对9具尸体标本进行4次尸体操作,评估Versius手术系统在机器人辅助胸外科手术中的性能。22/24(91.7%)胸科手术成功完成,包括17/18肺叶切除术、2/3胸腺切除术和3/3膈肌折叠术,端口和床旁单元配置验证成功,主刀医生认为该设备适合胸外科手术。

要点

  • 22/24(91.7%)胸科手术成功完成,包括17/18肺叶切除术、2/3胸腺切除术和3/3膈肌折叠术。
  • 1例胸腺切除术因尸体解剖原因未完成,1例肺叶切除术因控制台系统故障未完成。
  • 所有手术的端口和床旁单元配置均验证成功,主刀医生认为设备适合胸外科手术。

核心结论

  • 这项临床前研究证明该设备在尸体模型的机器人辅助胸外科手术中成功使用,并支持作为Idea、Development、Exploration、Assessment、Long-Term Follow Up-Devices第2a阶段(Development)的一部分推进至小规模临床研究。

研究设计与样本

  • 2018年11月至2020年12月间进行了4次尸体操作,评估设备在一系列胸科手术中的性能,手术分类为完成或未完成,外科医生评估设备成功完成必要手术步骤的能力,并记录端口和床旁单元位置。

主要发现

  • 在9具尸体标本中,共22/24(91.7%)胸科手术成功完成,包括17/18肺叶切除术、2/3胸腺切除术和3/3膈肌折叠术。
  • 1例胸腺切除术因尸体解剖原因未完成,1例肺叶切除术因控制台系统故障未完成。
  • 所有手术的端口和床旁单元配置均验证成功,主刀医生认为该设备非常适合胸外科手术。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The adoption of robot-assisted thoracic surgery (RATS) has helped to overcome some of the challenges associated with surgeons performing conventional video-assisted thoracic surgery. The Versius Surgical System (CMR Surgical, Cambridge, UK) has been developed iteratively in line with surgical team feedback to improve the surgeon's experience and patient outcomes. The goal of this study was to assess the use of the device in RATS in a preclinical setting and to fulfil Idea, Development, Exploration, Assessment, Long-Term Follow Up-Devices stage 1 (Idea). METHODS: Four cadaveric sessions were conducted between November 2018 and December 2020, during which device performance in a range of thoracic operations was assessed. Procedures were categorized as either completed or not completed, and surgeons evaluated the device's ability to successfully complete necessary surgical steps. Port and bedside unit positions were recorded. RESULTS: In total, 22/24 (91.7%) thoracic procedures were successfully completed, including 17/18 lobectomies, 2/3 thymectomies and 3/3 diaphragm plications, in 9 cadaver specimens. One thymectomy could not be completed due to cadaver anatomy and 1 lobectomy was not completed due a console system fault. Port and bedside unit configurations were successfully validated for all procedures, and lead surgeons deemed the device to be well-suited for thoracic surgery. CONCLUSIONS: This preclinical study demonstrated the successful use of the device in RATS in cadaveric models and supports progression to small-scale clinical studies, as part of Idea, Development, Exploration, Assessment, Long-Term Follow Up-Devices stage 2a (Development).