文献速递瓣膜外科

8字主动脉瓣缝合技术优化小主动脉瓣环的有效瓣口面积——一项离体研究

The figure-of-8 aortic valve suturing technique optimizes the effective orifice area of a small aortic annulus-an ex vivo study.
PubMed 文献 · Eur J Cardiothorac Surg · 2022-01-01

该离体研究在31个猪主动脉根部比较4种植入技术(非外翻带垫片缝合、单针间断缝合、连续缝合、8字缝合)及瓣环扩大后植入大一号瓣膜对血流动力学的影响。结果显示8字缝合与瓣环扩大后植入大一号瓣膜的有效瓣口面积相当,均优于非外翻带垫片缝合,且8字缝合未增加漏流量。

要点

  • 8字缝合技术联合21 mm瓣膜可获得与瓣环扩大后植入23 mm瓣膜相似的有效瓣口面积。
  • 单针间断缝合和连续缝合未较非外翻带垫片缝合增大有效瓣口面积。
  • 单针间断缝合的漏流量显著大于非外翻带垫片缝合,其余技术间无显著差异。

核心结论

  • 8字缝合技术联合21 mm瓣膜可获得与瓣环扩大后采用非外翻带垫片缝合植入23 mm瓣膜相似的有效瓣口面积,提示8字缝合可能是改善小瓣环血流动力学结果的有效手术改良,且无需额外瓣环扩大。

研究设计与样本

  • 离体研究,在31个猪主动脉根部使用21 mm PERIMOUNT Magna Ease®瓣膜,分别采用非外翻带垫片缝合、单针间断缝合、连续缝合、8字缝合4种技术,以及采用Nunez法瓣环扩大后以非外翻带垫片缝合植入23 mm瓣膜。
  • 使用模拟循环回路按ISO规定评估有效瓣口面积、平均压力阶差和漏流量。

主要发现

  • 每搏量74 ml时,21 mm瓣膜8字缝合的有效瓣口面积为1.81(1.70-1.85)cm²,大于非外翻带垫片缝合的1.68(1.63-1.72)cm²(P=0.005)。
  • 每搏量74 ml时,瓣环扩大后23 mm瓣膜的有效瓣口面积为1.83(1.73-1.92)cm²,大于21 mm瓣膜非外翻带垫片缝合(P<0.001)、单针间断缝合1.76(1.68-1.81)cm²(P=0.17)、连续缝合1.76(1.65-1.79)cm²(P=0.14),但与8字缝合无显著差异。
  • 单针间断缝合和连续缝合较非外翻带垫片缝合未获得更大的有效瓣口面积。
  • 每搏量74 ml时,21 mm瓣膜单针间断缝合的漏流量为6.00(4.02-7.06)ml/次,大于非外翻带垫片缝合的3.51(1.85-4.53)ml/次(P<0.001);连续缝合4.04(3.60-4.49)ml/次(P=0.10)、8字缝合3.16(1.99-3.62)ml/次(P=0.74)与非外翻带垫片缝合无显著差异;瓣环扩大后23 mm瓣膜非外翻带垫片缝合为2.89(2.45-4.72)ml/次(P=0.51)。

临床怎么用

  • 作者认为8字缝合技术可能是改善小瓣环血流动力学结果的有效手术改良,且无需额外瓣环扩大。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Surgical aortic valve replacement (SAVR) in small annuli carries an elevated risk for the patient-prosthesis mismatch. In this study, we systematically investigated the influence of different implantation techniques including annular enlargement (AE) on the functional result after SAVR in small annuli using a standardized ex vivo model. METHODS: SAVR using the PERIMOUNT Magna Ease® (PME) 21 mm was performed in small porcine aortic roots using 4 implantation techniques: non-everting pledgeted (NE) suture, single interrupted (SI) suture, continuous suture (CS), figure-of-8 (F8) suture, as well as the PME 23 mm after AE using the Nunez method and the NE suture technique (PME23 AE). The effective orifice area (EOA), mean pressure gradient and leakage volume were evaluated using a mock circulation loop in accordance with ISO regulations. RESULTS: Experiments were conducted on 31 porcine aortic roots. PME21 using F8 and PME23 after AE achieved a significantly larger EOA than using NE. PME23 after AE showed a larger EOA than the PME21 using any suture technique, except the F8 [for stroke volume of 74 ml: PME21 NE: 1.68 (1.63-1.72) cm2, PME21 SI: 1.76 (1.68-1.81) cm2 (P = 0.17), PME21 CS: 1.76 (1.65-1.79) cm2 (P = 0.14), PME21 F8: 1.81 (1.70-1.85) cm2 (P = 0.005); PME23 AE: 1.83 (1.73-1.92) cm2 (P < 0.001)]. SI and CS did not result in larger EOA compared with the NE technique. PME21 using SI had a significantly larger leakage volume than using NE and there was no significant difference between other techniques [for stroke volume of 74 ml: PME21 NE: 3.51 (1.85-4.53) ml/stroke, PME21 SI: 6.00 (4.02-7.06) ml/stroke (P < 0.001), PME21 CS: 4.04 (3.60-4.49) ml/stroke (P = 0.10), PME21 F8: 3.16 (1.99-3.62) ml/stroke (P = 0.74), PME23 NE: 2.89 (2.45-4.72) ml/stroke (P = 0.51)]. CONCLUSIONS: The F8 technique with the PME21 achieved a similar EOA as the 1 size larger PME23 using NE after AE. These results suggest that the F8 technique may be an effective surgical modification to improve the haemodynamic result in a small annulus without additional AE.

来源:Eur J Cardiothorac Surg,查看原文。版权归原作者所有。

引用

  1. Saisho H, Kuhlmann C, Scharfschwerdt M, et al. The figure-of-8 aortic valve suturing technique optimizes the effective orifice area of a small aortic annulus-an ex vivo study. Eur J Cardiothorac Surg. 2022. PMID: 35380636. doi:10.1093/ejcts/ezac195