Castor单分支支架治疗累及左锁骨下动脉的穿透性主动脉溃疡的胸主动脉腔内修复结局
研究纳入2019年1月至2020年12月共25例需左锁骨下动脉血运重建的穿透性主动脉溃疡患者,行Castor单分支支架胸主动脉腔内修复。24例(96.0%)手术成功,1例(4.0%)I型内漏、1例(4.0%)卒中、1例(4.0%)支架移位、1例(4.0%)截瘫;平均随访15.7个月,2例分支闭塞、1例心肌梗死相关死亡。结论认为初步结果支持该装置的有效性和安全性。
要点
- 25例穿透性主动脉溃疡患者使用Castor单分支支架行胸主动脉腔内修复,24例(96.0%)手术成功。
- 围术期并发症包括I型内漏1例(4.0%)、卒中1例(4.0%)、支架移位1例(4.0%)、截瘫1例(4.0%)。
- 平均随访15.7个月,发生分支闭塞2例、心肌梗死相关死亡1例。
核心结论
- 这些初步结果支持Castor单分支支架在治疗需左锁骨下动脉血运重建的穿透性主动脉溃疡中的有效性和安全性。
研究设计与样本
- 研究纳入2019年1月至2020年12月共25例穿透性主动脉溃疡患者,均需在胸主动脉腔内修复中使用Castor装置行左锁骨下动脉血运重建。
- 分析指标包括手术结果、30天及随访死亡与并发症、主动脉计算机断层扫描血管造影数据。
主要发现
- 24例(96.0%)手术成功,1例(4.0%)发生I型内漏,1例(4.0%)发生卒中。
- 1例(4.0%)发生支架移位,1例(4.0%)发生截瘫,2例(8.0%)置入两枚支架。
- 平均随访15.7个月,发现2例分支闭塞和1例心肌梗死相关死亡。
- 主动脉计算机断层扫描血管造影显示,在12个月随访期内溃疡直径和深度显著减小,所有病例溃疡腔完全血栓形成。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTS: To acquire adequate proximal landing zones for penetrating aortic ulcers in the proximal descending aorta or aortic arch, the left subclavian artery (LSA) is usually covered during thoracic endovascular aortic repair. The Castor single-branched stent graft is used to treat aortic disease involving LSA. Our goal was to evaluate the efficiency and safety of the Castor device in treating penetrating aortic ulcers (PAUs) involving the LSA. METHODS: Between January 2019 and December 2020, we enrolled 25 patients with PAUs requiring LSA revascularization using Castor devices during thoracic endovascular aortic repair (TEVAR). Outcomes including surgical results, 30-day and follow-up deaths and morbidity and aortic computed tomography angiography data were analyzed. RESULTS: In total, 24 patients (96.0%) underwent successful surgical procedures: 1 patient (4.0%) had a type I endoleak and 1 patient (4.0%) had a stroke. Stent migration occurred in 1 patient (4.0%) and paraplegia in 1 patient (4.0%). Two stents were deployed in 2 patients (8.0%). After a mean follow-up period of 15.7 months, 2 cases of branch occlusions and 1 death related to myocardial infarction were noted. Aortic computed tomography angiography demonstrated that the diameters and depths were reduced significantly whereas complete thrombosis of the lumens of the PAUs occurred in all cases during the 12-months follow-up period. CONCLUSIONS: These preliminary results justify the efficiency and safety of the Castor single branched stent graft in the treatment of PAUs requiring LSA revascularization.
引用
- Huang B, Jia H, Lai H, et al. Outcomes of thoracic endovascular aortic repair for penetrating aortic ulcers involving the left subclavian artery with the Castor single-branched stent graft. Eur J Cardiothorac Surg. 2022. PMID: 35274139. doi:10.1093/ejcts/ezac102
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。