HVAD更换为HeartMate 3左心室辅助装置后的移植物阻力差异

Graft Resistance Difference after HVAD to HeartMate 3 Left Ventricular Assist Device Exchange.
PubMed 文献 · Ann Thorac Surg · 2022-01-03

该体外稳态研究对比了HVAD、HeartMate 3及混合移植物在3.0至6.0 L/min流量下的阻力差异。在4.3和6.0 L/min流量下,使用全HVAD移植物时HeartMate 3转子转速需分别从5350升至5700 rpm和从6350升至6900 rpm,功耗分别增加16%和24%。作者认为完全更换移植物仍是首选方案。

要点

  • 在4.3 L/min流量下,全HVAD移植物使HeartMate 3转速从5350升至5700 rpm,功耗增加16%。
  • 在6.0 L/min流量下,全HVAD移植物使转速从6350升至6900 rpm,功耗增加24%。
  • 作者认为完全更换为HeartMate 3移植物仍是工程学首选方案。

核心结论

  • 研究未阐明泵更换时使用残留HVAD移植物的严重后果,但需权衡更高转速的增量风险、电池续航劣势与完全转换至HeartMate 3移植物的总体益处。
  • 从工程学角度,HVAD转换为HeartMate 3时完全更换移植物仍是首选方案。

研究设计与样本

  • 采用稳态体外研究,使用250 mm长的HVAD、HeartMate 3及半HVAD半HeartMate 3移植物样本,另使用125 mm和375 mm长的HVAD移植物样本。
  • 流量设置为3.0、3.9、4.3、4.7和6.0 L/min以涵盖预期临床条件。

主要发现

  • 在4.3 L/min流量下,全HVAD移植物的HeartMate 3转子转速为5700 rpm,而HeartMate 3移植物为5350 rpm。
  • 在6.0 L/min流量下,全HVAD移植物的HeartMate 3转子转速为6900 rpm,而HeartMate 3移植物为6350 rpm。
  • 在4.3 L/min流量下,全HVAD移植物的功耗为4.3 W,较HeartMate 3移植物的3.7 W增加16%。
  • 在6.0 L/min流量下,全HVAD移植物的功耗为6.8 W,较HeartMate 3移植物的5.5 W增加24%。

临床怎么用

  • 作者认为,从工程学角度,HVAD转换为HeartMate 3时完全更换移植物仍是首选方案。
  • 作者指出,需将更高转速的增量风险、电池续航劣势与完全转换至HeartMate 3移植物的总体益处同其他操作考虑进行权衡。

局限性

  • 研究为体外稳态研究,未阐明使用残留HVAD移植物时的严重后果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: A likely consequence of the discontinued distribution and sale of the HVAD System (Medtronic, Minneapolis, MN) will be an increase in replacement with the HeartMate 3 (Abbott, Chicago, IL) left ventricular assist device when device exchange is necessary. If part or all of the HVAD 10-mm-diameter outflow graft is retained during replacement, the HeartMate 3 will have to run at a higher speed than it would with its 14-mm-diameter graft. METHODS: A steady-state, in vitro study was run with 250-mm-long samples of HVAD, HeartMate 3, and half-HVAD/half-HeartMate 3 grafts and additionally 125- and 375-mm-long samples of the HVAD graft. Flows of 3.0, 3.9, 4.3, 4.7, and 6.0 L/min were applied to encompass expected clinical conditions. RESULTS: At typical and high flow rates of 4.3 and 6.0 L/min, HeartMate 3 rotor speeds with the full HVAD graft had to be increased relative to those with the HeartMate 3 graft from 5350 to 5700 and 6350 to 6900 rpm, respectively, with power consumption increases from 3.7 to 4.3 W (16%) and 5.5 to 6.8 W (24%), respectively. CONCLUSIONS: The study did not elucidate a severe consequence of using a remnant HVAD graft during pump exchange, but the incremental risks of a higher rotor speed, disadvantage to the patient in battery runtime, and the general benefit of complete conversion to the HeartMate 3 graft should be balanced against other procedural considerations. Complete graft replacement during HVAD-to-HeartMate 3 conversion remains the preferred approach from an engineering point of view.