免缝合主动脉瓣生物假体支架变形:基于影像与三维建模的先导观察性分析
该先导观察性分析纳入2014年8月至2019年5月共134例接受Perceval免缝合主动脉瓣生物假体置换的患者,对其中16例用胸部CT重建支架三维模型并分析椭圆度,同时测量径向力。结果显示所有重建支架均有不同程度变形和椭圆化,径向力明显低于Edwards Intuity及多种经导管主动脉瓣植入装置。
要点
- 全部16例三维重建的Perceval支架均出现不同程度变形与椭圆化。
- 瓣环与交界区椭圆度范围分别为6.8%~45%,均值13.6%与21.9%。
- Perceval径向力明显低于Edwards Intuity及多种经导管主动脉瓣植入装置。
核心结论
- 所有分析的Perceval生物假体均出现不同程度变形,相对较低的径向力可能是其变形倾向的原因,变形可导致瓣叶扑动和活动度下降,并可能引起纤维化、跨瓣压差升高,以及乳酸脱氢酶升高和血小板计数下降。
研究设计与样本
- 先导观察性分析,2014年8月至2019年5月共134例患者(52%女性)接受Perceval生物假体主动脉瓣置换。
- 对其中16例使用Mimics软件基于胸部CT扫描重建Perceval支架三维模型并分析支架椭圆度。
- 进行径向力测量,比较Perceval生物假体与其他瓣膜的刚度。
主要发现
- 三维重建显示所有Perceval支架均存在一定程度变形和椭圆化。
- Perceval支架瓣环区和交界区椭圆度范围为6.8%~45%,均值分别为13.6%和21.9%。
- Perceval假体径向力明显低于Edwards Intuity及多种经导管主动脉瓣植入装置。
- 在2例有术前和术后CT数据的患者中,支架采用了主动脉的术前形态。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The goal of this analysis of the Perceval aortic bioprosthesis was to investigate the ovalization/deformation of the Perceval prosthesis stent after implantation and its propensity for cusp dysfunction, fluttering, fibrosis and blockage. METHODS: Between August 2014 and May 2019, a total of 134 patients (52% female) underwent aortic valve replacement with the Perceval bioprosthesis. We reconstructed three-dimensional models of the Perceval stent using thorax computed tomography scans for 16 patients employing the software Mimics (Materialise NV, Leuven, Belgium) and analysed the ovality of the stents. Radial force (RF) measurements were performed to compare the stiffness of the Perceval bioprosthesis to that of other valves. RESULTS: The three-dimensional reconstructions showed that all Perceval stents exhibited some degree of deformation and ovalization. Ovality in the annulus and commissure section of the Perceval stents ranged from 6.8% to 45% with mean values of 13.6% and 21.9%, respectively. The RF of the Perceval prosthesis was noticeably lower than that of the Edwards Intuity and several transcatheter aortic valve implantation devices. The stent adopted the preoperative shape of the aorta in the 2 patients for whom pre- and postoperative computed tomography data existed. CONCLUSIONS: The Perceval bioprostheses were deformed to different degrees in all analysed cases. The comparably low RFs might be an explanation for the propensity towards this deformation, which can lead to fluttering, a reduction of the cusps' mobility. This condition could potentially result in fibrosis as well as increased transvalvular pressure gradients and might be the cause for the increase in lactate dehydrogenase and the decrease in platelet count.
引用
- Aljalloud A, Spetsotaki K, Tewarie L, et al. Stent deformation in a sutureless aortic valve bioprosthesis: a pilot observational analysis using imaging and three-dimensional modelling. Eur J Cardiothorac Surg. 2022. PMID: 34791144. doi:10.1093/ejcts/ezab485
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。