文献速递产科麻醉区域麻醉与超声

可生物吸收封堵装置在羊模型中封堵Tuohy针硬膜穿刺后渗漏的验证

Validation of a bioabsorbable device that seals perforations after Tuohy needle dural puncture in an ovine model.
PubMed 文献 · Reg Anesth Pain Med · 2021-01-27

研究设计了一种经硬膜外穿刺针直接封堵意外硬膜穿刺的装置,在46只羊中单盲随机对照验证。结果显示封堵组硬膜外腔渗漏量中位数1.0 mL低于对照组1.4 mL(p=0.008),硬膜外/感兴趣区比值中位数29低于46(p=0.013),表明该装置可减少脑脊液渗漏。

要点

  • 该封堵装置可减少硬膜穿刺后脑脊液向硬膜外腔的渗漏量。
  • 封堵组硬膜外渗漏量中位数1.0 mL,对照组1.4 mL(p=0.008)。
  • 硬膜外/ROI比值中位数封堵组29,对照组46(p=0.013)。

核心结论

  • 新型硬膜穿刺封堵装置可减少硬膜穿刺后脑脊液向硬膜外腔的渗漏量。

研究设计与样本

  • 单盲随机对照动物实验,46只麻醉成年羊等分为两组,均在L6-S1间以18G Tuohy针行硬膜穿刺。

主要发现

  • 封堵组硬膜外渗漏量中位数1.0 mL,对照组1.4 mL(p=0.008)。
  • 封堵组硬膜外/ROI比值中位数29,对照组46(p=0.013;95%CI -27至-3.5)。

临床怎么用

  • 该装置可在识别硬膜穿刺后、拔出穿刺针前经针直接封堵,可能用于减少即刻脑脊液渗漏。

局限性

  • 研究为羊模型,样本量46只,装置在2例中因制造缺陷未能成功部署;临床适用性需进一步验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: We designed a device to close accidental dural puncture via the offending puncturing epidural needle directly after diagnosis of the puncture and before removing the needle. The aim of this study was to quantify this device's ability to seal cerebrospinal fluid leakage. METHODS: Forty-six anesthetized adult sheep were studied in a single-blind randomized controlled fashion in two equal groups.An intentional dural puncture was performed with an 18-gage Tuohy needle on all the sheep between L6 and S1 levels. Contrast medium was injected through the needle. Twenty-three animals receive treatment with the sealing device. Two minutes after device placement, or dural puncture in the control group, a CT scan was performed on the animals to estimate contrast material leakage. A region of interest (ROI) was defined as the region that enclosed the subarachnoid space, epidural space, and neuroforaminal canal (the vertebral body above and half of its equivalent height in sacrum below the puncture site). In this region, the total contrast volume and the volumes in the epidural space (EPIDURAL) were measured. The primary outcome measure was the EPIDURAL/ROI ratio to ascertain the proportion of intrathecally injected fluid that passed into the epidural space in both groups. The secondary outcomes were the total amount of contrast in the ROI and the EPIDURAL. RESULTS: The device was deployed successfully in all but two instances, where it suffered from manufacturing defects.Leakage was less in the study group (1.0 vs 1.4 mL, p=0.008). The median EPIDURAL/ROI ratio was likewise less in the study group (29 vs 46; p=0.013; 95% CI (-27 to -3.5)). CONCLUSION: This novel dural puncture-sealing device, also envisaged to be used in other comparable iatrogenic leakage scenarios to be identified in the future, was able to reduce the volume of cerebrospinal fluid that leaked into the epidural space after dural puncture. The device is possibly a valuable way of preventing fluid leakage immediately after the recognition of membrane puncture.

来源:Reg Anesth Pain Med,查看原文。版权归原作者所有。

引用

  1. García-Vitoria C, García-Roselló M, Reina MA, et al. Validation of a bioabsorbable device that seals perforations after Tuohy needle dural puncture in an ovine model. Reg Anesth Pain Med. 2021. PMID: 33504475. doi:10.1136/rapm-2020-102225

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。