文献速递区域麻醉与超声

针尖追踪与可视化新方法

Novel approaches to needle tracking and visualisation.
PubMed 文献 · Anaesthesia · 2021-01-01

超声引导神经阻滞的准确性和可靠性仍不足,针尖与神经外膜关系难以评估。创新转向利用组织物理特性(压力、电、光学、声学、弹性)的生物标志物检测针尖位置。多数处于实验室阶段,仅phantom或尸体研究结果,临床试验即将开展。

要点

  • 超声分辨率有限且进针角度大时显影困难
  • 新型针尖追踪技术利用压力、阻抗、光学等物理特性
  • 多数技术尚处实验室阶段,临床试验即将开展

核心结论

  • 超声引导神经阻滞的准确性和可靠性尚不足以保障所有患者的安全阻滞。
  • 创新方向转向利用组织物理特性的生物标志物检测针尖位置。

研究设计与样本

  • 综述性文章,总结针尖追踪与可视化新技术。
  • 提及的技术多处于实验室阶段,结果仅来自phantom或尸体研究。

主要发现

  • 超声分辨率不足,组织因声阻抗难以辨别,进针角度增大时针显影更困难。
  • 置于阻滞针腔内的细光纤可测量针尖压力。
  • 电阻抗可区分针尖位于神经内或神经周围。
  • 新型针尖追踪针远端嵌入压电元件,以蓝/红或绿圈在平面内和平面外追踪针尖。
  • 针尖微型超声正在开发,40MHz在麻醉猪中显示肌纹、清晰神经外膜及30-40个直径>75微米的束。

临床怎么用

  • 当前临床仍应认识到超声引导神经阻滞的局限性。
  • 关注针尖追踪新技术进展,未来可能改善患者安全并改变实践。

局限性

  • 多数技术仅处于实验室阶段,结果仅来自phantom或尸体研究。
  • 临床试验尚未开展,临床有效性和安全性未证实。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The accuracy and reliability of ultrasound are still insufficient to guarantee complete and safe nerve block for all patients. Injection of local anaesthetic close to, but not touching, the nerve is key to outcomes, but the exact relationship between the needle tip and nerve epineurium is difficult to evaluate, even with ultrasound. Ultrasound has insufficient resolution, tissues are difficult to discern due to acoustic impedance and needles are more difficult to see with increased angulation. The limitations of ultrasound have shifted the focus of innovation towards bio-markers that help detect needle tip position by utilising the physical properties of tissues, (e.g. pressure, electrical, optics, acoustic and elastic). Although most are at the laboratory stage and results are as yet only available from phantom or cadaver studies, clinical trials are imminent. For example, fine optical fibres placed within the lumen of block needles can measure needle tip pressure. Electrical impedance differentiates between intraneural and perineural needle tip placement. A new tip tracker needle has a piezo element embedded at its distal end that tracks the needle tip in-plane and out-of-plane as a blue/red or green circle depending on its relative location within the beam. Micro-ultrasound at the tip of the needle is in development. Early images using 40MHz in anaesthetised pigs reveal muscle striation, distinct epineurium and 30-40 fascicles > 75 micron in diameter. The next few years will see a technological revolution in tip-tracking technology that has the potential to improve patient safety and, in doing so, change practice.

来源:Anaesthesia,查看原文。版权归原作者所有。

引用

  1. McLeod GA. Novel approaches to needle tracking and visualisation. Anaesthesia. 2021. PMID: 33426657. doi:10.1111/anae.15232