麻醉 adequaci 监测与标准临床监测在常规全麻中的比较:国际多中心单盲随机对照试验

Comparison of adequacy of anaesthesia monitoring with standard clinical practice monitoring during routine general anaesthesia: An international, multicentre, single-blinded randomised controlled trial.
PubMed 文献 · Eur J Anaesthesiol · 2021-01-01

该研究为国际多中心单盲随机对照试验,纳入494例择期气管插管成年手术患者,比较Entropy联合手术体积描记指数指导麻醉与标准监测对麻醉不adequaci事件的影响。两组不adequaci事件发生率约0.7次/患者且无差异(P=0.519),但AoA组丙泊酚用量显著减少(6.9 vs. 7.5 mg kg h,P=0.008),苏醒时间(8.0 vs. 9.6 min,P=0.005)及PACU完全恢复时间更短(P=0.043)。

要点

  • AoA监测未减少麻醉不adequaci事件(约0.7次/患者,P=0.519)。
  • AoA组丙泊酚用量显著降低(6.9 vs. 7.5 mg kg h,P=0.008)。
  • AoA组苏醒时间(8.0 vs. 9.6 min,P=0.005)和PACU恢复更短。

核心结论

  • 以Entropy和手术体积描记指数指导全凭静脉麻醉,与标准监测相比未能减少不adequaci麻醉事件,但减少了丙泊酚用量并缩短了苏醒和PACU停留时间。

主要发现

  • 两组不adequaci麻醉事件发生率约0.7次/患者,组间无差异(P=0.519)。
  • AoA组丙泊酚总消耗量显著低于对照组(6.9 vs. 7.5 mg kg h,P=0.008),瑞芬太尼消耗量两组相等。
  • AoA组苏醒时间(8.0 vs. 9.6 min,P=0.005)及PACU完全恢复时间显著更短(P=0.043)。
  • 术后疼痛评分及镇痛药使用两组无差异。

临床怎么用

  • 在丙泊酚-瑞芬太尼靶控输注全麻中,可考虑使用Entropy和手术体积描记指数监测以减少丙泊酚用量并加速苏醒,但不应期望降低不adequaci麻醉事件。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Previous studies have suggested that monitoring the levels of both hypnosis and antinociception could reduce periods of inadequate anaesthesia. However, the evidence regarding associated benefits of this monitoring is still limited. OBJECTIVE: The primary objective of this study was to confirm that guidance of anaesthesia by depth of hypnosis and antinociception monitoring decreases the number of inadequate anaesthesia events in comparison with standard clinical practice. DESIGN: A multicentre, single-blinded, randomised controlled trial. SETTING: The study was conducted in four European University hospitals in four different countries between December 2013 and November 2016. PATIENTS: The study population consisted of a total of 494 adult patients undergoing elective surgery requiring tracheal intubation. INTERVENTIONS: The patients were allocated to one of two groups. The first group was treated using Entropy for depth of hypnosis and surgical pleth index to determine depth of antinociception (adequacy of anaesthesia group; AoA group). The second group was monitored using standard monitoring alone (control group). Anaesthesia was conducted with target-controlled infusions of propofol and remifentanil. MAIN OUTCOME MEASURES: The primary outcome of the study was the number of total unwanted events for example signs of inadequately light or unintentionally deep anaesthesia. RESULTS: Evidence of inadequate anaesthesia had an incidence of around 0.7 events per patient in both groups with no difference between groups (P = 0.519). In the AoA group, the overall consumption of propofol was significantly reduced (6.9 vs. 7.5 mg kg h, P = 0.008) in comparison with the control group. The consumption of remifentanil was equal in both groups. The times to emergence [8.0 vs. 9.6 min (P = 0.005)] and full recovery in the postanaesthesia care unit (P = 0.043) were significantly shorter in the AoA group. No differences were seen in postoperative pain scores or in the use of analgesics. CONCLUSION: In the current study, the guidance of total intravenous anaesthesia by Entropy and surgical pleth index in comparison with standard monitoring alone was not able to validate reduction of unwanted anaesthesia events. However, there was a reduction in the use of propofol, and shorter times for emergence and time spent in the postanaesthesia care unit. TRIAL REGISTRATION: at ClinicalTrials.gov NCT01928875.

来源:Eur J Anaesthesiol,查看原文。版权归原作者所有。

引用

  1. Gruenewald M, Harju J, Preckel B, et al. Comparison of adequacy of anaesthesia monitoring with standard clinical practice monitoring during routine general anaesthesia: An international, multicentre, single-blinded randomised controlled trial. Eur J Anaesthesiol. 2021. PMID: 33074943. doi:10.1097/EJA.0000000000001357