气管插管与拔管气溶胶生成的定量评估
在超洁净手术室实时监测气管插管和拔管期间气溶胶颗粒。插管(含面罩通气)产气溶胶极低(1.4颗粒/升),拔管为21颗粒/升,均低于自主咳嗽的732颗粒/升。研究不支持将择期插管列为气溶胶生成操作。
要点
- 择期气管插管(含面罩通气)产气溶胶极低,不支持列为气溶胶生成操作。
- 拔管产气溶胶高于插管但远低于自主咳嗽,未达高风险气溶胶生成操作标准。
- 需重新评估常规气道管理的气溶胶生成操作定义及防护措施。
核心结论
- 择期气管插管不应被列为气溶胶生成操作。
- 拔管产气溶胶多于插管但低于高风险气溶胶生成操作标准。
研究设计与样本
- 在超洁净通风手术室进行实时高分辨率环境监测。
- 使用光学粒子计数器连续采样,监测患者与麻醉医生之间区域。
- 插管组n=14,拔管组n=10,自主咳嗽阳性对照n=38。
主要发现
- 背景颗粒计数极低(0.4颗粒/升),可分辨气道管理相关的瞬态颗粒增加。
- 自主咳嗽平均浓度732(418)颗粒/升(n=38)。
- 插管含面罩通气平均浓度1.4(1.4)颗粒/升(n=14,p<0.0001 vs. 咳嗽)。
- 拔管平均浓度21(18)颗粒/升(n=10),为插管的15倍(p=0.0004),为自主咳嗽的1/35(p<0.0001)。
临床怎么用
- 为气道管理气溶胶风险评估提供定量方法,可扩展至其他技术。
- 提示需重新评估气溶胶生成操作定义及常规麻醉气道管理的防护措施。
局限性
- 研究在超洁净通风手术室进行,结果外推至其他环境需谨慎。
- 样本量较小(插管n=14,拔管n=10)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The potential aerosolised transmission of severe acute respiratory syndrome coronavirus-2 is of global concern. Airborne precaution personal protective equipment and preventative measures are universally mandated for medical procedures deemed to be aerosol generating. The implementation of these measures is having a huge impact on healthcare provision. There is currently a lack of quantitative evidence on the number and size of airborne particles produced during aerosol-generating procedures to inform risk assessments. To address this evidence gap, we conducted real-time, high-resolution environmental monitoring in ultraclean ventilation operating theatres during tracheal intubation and extubation sequences. Continuous sampling with an optical particle sizer allowed characterisation of aerosol generation within the zone between the patient and anaesthetist. Aerosol monitoring showed a very low background particle count (0.4 particles.l-1 ) allowing resolution of transient increases in airborne particles associated with airway management. As a positive reference control, we quantitated the aerosol produced in the same setting by a volitional cough (average concentration, 732 (418) particles.l-1 , n = 38). Tracheal intubation including facemask ventilation produced very low quantities of aerosolised particles (average concentration, 1.4 (1.4) particles.l-1 , n = 14, p < 0.0001 vs. cough). Tracheal extubation, particularly when the patient coughed, produced a detectable aerosol (21 (18) l-1 , n = 10) which was 15-fold greater than intubation (p = 0.0004) but 35-fold less than a volitional cough (p < 0.0001). The study does not support the designation of elective tracheal intubation as an aerosol-generating procedure. Extubation generates more detectable aerosol than intubation but falls below the current criterion for designation as a high-risk aerosol-generating procedure. These novel findings from real-time aerosol detection in a routine healthcare setting provide a quantitative methodology for risk assessment that can be extended to other airway management techniques and clinical settings. They also indicate the need for reappraisal of what constitutes an aerosol-generating procedure and the associated precautions for routine anaesthetic airway management.
引用
- Brown J, Gregson FKA, Shrimpton A, et al. A quantitative evaluation of aerosol generation during tracheal intubation and extubation. Anaesthesia. 2020. PMID: 33022093. doi:10.1111/anae.15292